Open official notice

Stryker harmonic focus shears + adaptive tissue technology (reprocessed)

Official notice

26 Sept 2025

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Stryker harmonic focus shears + adaptive tissue technology (reprocessed). Stryker has received an increase in complaints for HAR9F continuous activation due to the min button becoming stuck in the on position. Stryker has investigated and identified the manufacturing issue which contributed to the continuous activation noted with these devices. As of August 31, 2025, device failures related to this issue represent 0.16% of distributed devices; however, there have been no reported serious injuries. Published 2025-09-26.

What to do now

Contact the manufacturer if you require additional information.

Recall number

78150

Official notice

2025-09-26

The hazard

Stryker has received an increase in complaints for HAR9F continuous activation due to the min button becoming stuck in the on position. Stryker has investigated and identified the manufacturing issue which contributed to the continuous activation noted with these devices. As of August 31, 2025, device failures related to this issue represent 0.16% of distributed devices; however, there have been no reported serious injuries.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

78150

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - Stryker Harmonic Focus Shears + Adaptive Tissue Technology (Reprocessed), Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - HAR9F

Hazard

Stryker has received an increase in complaints for HAR9F continuous activation due to the min button becoming stuck in the on position. Stryker has investigated and identified the manufacturing issue which contributed to the continuous activation noted with these devices. As of August 31, 2025, device failures related to this issue represent 0.16% of distributed devices; however, there have been no reported serious injuries.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Stryker Sustainability, Stryker Harmonic Focus Shears + Adaptive Tissue Technology.

Source & status

Tier

official

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

78150

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 26 Sept 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button