Image not available in RecallScope.
Stryker harmonic focus shears + adaptive tissue technology (reprocessed)
Official notice
26 Sept 2025
An official Health Canada notice concerns Stryker harmonic focus shears + adaptive tissue technology (reprocessed). Stryker has received an increase in complaints for HAR9F continuous activation due to the min button becoming stuck in the on position. Stryker has investigated and identified the manufacturing issue which contributed to the continuous activation noted with these devices. As of August 31, 2025, device failures related to this issue represent 0.16% of distributed devices; however, there have been no reported serious injuries. Published 2025-09-26.
What to do now
Contact the manufacturer if you require additional information.
Recall number
78150
Official notice
2025-09-26
The hazard
The hazard
Stryker has received an increase in complaints for HAR9F continuous activation due to the min button becoming stuck in the on position. Stryker has investigated and identified the manufacturing issue which contributed to the continuous activation noted with these devices. As of August 31, 2025, device failures related to this issue represent 0.16% of distributed devices; however, there have been no reported serious injuries.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Stryker Sustainability, Stryker Harmonic Focus Shears + Adaptive Tissue Technology.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.