Novum iq lvp
Official notice
22 Jul 2025
Last checked
22 Jul 2025
An official Health Canada notice concerns Novum iq lvp. Baxter is issuing an urgent medical device correction for the Novum IQ Large Volume Pump (LVP) due to the following two issues: there is a potential for underinfusion when transitioning from a lower to a higher flow rate (e.g., rate change or bolus). Specifically, the risk occurs when the second flow rate is more than double the first flow rate. Baxter has identified an increase in customer reports of over and underinfusion that may be due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Recall start date: July 17, 2025 Published 2025-07-22.
What to do now
Contact the manufacturer if you require additional information.
Recall number
77756
Official notice
2025-07-22
The hazard
The hazard
Baxter is issuing an urgent medical device correction for the Novum IQ Large Volume Pump (LVP) due to the following two issues: there is a potential for underinfusion when transitioning from a lower to a higher flow rate (e.g., rate change or bolus). Specifically, the risk occurs when the second flow rate is more than double the first flow rate. Baxter has identified an increase in customer reports of over and underinfusion that may be due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Recall start date: July 17, 2025
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Baxter Healthcare, Novum Iq Lvp.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 22 Jul 2025. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.