Novum iq lvp

Official notice

22 Jul 2025

RecallScope risk estimate:MediumHealth Canada

Baxter is issuing an urgent medical device correction for the novum iq large volume pump (lvp) due to the following two issues: there is a potential for underinfusion when transitioning from a lower to a higher flow rate (e.g., rate change or bolus). Specifically, the risk occurs when the second flow rate is more than double the first flow rate. Baxter has identified an increase in customer reports of over and underinfusion that may be due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Recall start date: july 17, 2025

What to do now

Contact the manufacturer if you require additional information.

Recall number

77756

Official notice

2025-07-22

The hazard

Baxter is issuing an urgent medical device correction for the Novum IQ Large Volume Pump (LVP) due to the following two issues: there is a potential for underinfusion when transitioning from a lower to a higher flow rate (e.g., rate change or bolus). Specifically, the risk occurs when the second flow rate is more than double the first flow rate. Baxter has identified an increase in customer reports of over and underinfusion that may be due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Recall start date: July 17, 2025

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

77756

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - Novum IQ LVP, Lot or serial number - All lots., Model or catalogue - 40700BAX

Hazard

Baxter is issuing an urgent medical device correction for the Novum IQ Large Volume Pump (LVP) due to the following two issues: there is a potential for underinfusion when transitioning from a lower to a higher flow rate (e.g., rate change or bolus). Specifically, the risk occurs when the second flow rate is more than double the first flow rate. Baxter has identified an increase in customer reports of over and underinfusion that may be due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Recall start date: July 17, 2025

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Baxter Healthcare, Novum Iq Lvp.

Source & status

Tier

official

Category

Health

Official classification

Type II

Recall details

Recall number

77756

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 22 Jul 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button