Magna-Fx Cannulated Screw Fixation System 7.0 mm. Item codes starting with 0011420, 11460 and 11461

Official notice

26 Feb 2014

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Magna-Fx Cannulated Screw Fixation System 7.0 mm. Item codes starting with 0011420, 11460 and 11461. Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations. Published 2014-02-26.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

7,044,680

Recall number

Z-1001-2014

Official notice

2014-02-26

The hazard

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

66500

Recall Numbers

Z-1001-2014

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Magna-Fx Cannulated Screw Fixation System 7.0 mm. Item codes starting with 0011420, 11460 and 11461.

Units affected

7,044,680

Hazard

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

BE · Belgium

Affected area 3

CA · Canada

Affected area 4

CH · Switzerland

Affected area 5

CN · China

Affected area 6

CZ · Czech Republic

Affected area 7

DE · Germany

Affected area 8

DK · Denmark

Affected area 9

ES · Spain

Affected area 10

FI · Finland

Affected area 11

FR · France

Affected area 12

GB · United Kingdom

Affected area 13

GR · Greece

Affected area 14

IN · India

Affected area 15

IT · Italy

Affected area 16

JP · Japan

Affected area 17

KR · South Korea

Affected area 18

NL · Netherlands

Affected area 19

NO · Norway

Affected area 20

NZ · New Zealand

Affected area 21

PT · Portugal

Affected area 22

SE · Sweden

Affected area 23

SG · Singapore

Affected area 24

SK · Slovakia

Affected area 25

TH · Thailand

Affected area 26

TR · Turkey

Affected area 27

TW · Taiwan

Company & contacts

Brand

Manufacturer

Zimmer, Inc.

Phone

Not provided

Brand: Zimmer. Contact: Zimmer, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1001-2014

Event ID

66500

Units affected

7,044,680

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 26 Feb 2014. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button