Immulite 2000 system-human chorionic gonadotropin (hcg) assay

Official notice

28 Mar 2022

Last checked

28 Mar 2022

Medium riskHealth Canada
Watch Siemens

An official Health Canada notice concerns Immulite 2000 system-human chorionic gonadotropin (hcg) assay. Siemens Healthcare Diagnostics Inc. Has confirmed the potential for falsely elevated hcg results due to sample carryover. This can be observed when a sample is assayed for HCG immediately after an undiluted sample with a HCG value of GT;5000 MIU/mL. This issue impacts serum and urine patient samples, as well as quality control samples and adjustors. The effect of carryover varies based on the concentration of the high HCG sample. Recall start date: Mar 1, 2022 Published 2022-03-28.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64028

Official notice

2022-03-28

The hazard

Siemens Healthcare Diagnostics Inc. Has confirmed the potential for falsely elevated hcg results due to sample carryover. This can be observed when a sample is assayed for HCG immediately after an undiluted sample with a HCG value of GT;5000 MIU/mL. This issue impacts serum and urine patient samples, as well as quality control samples and adjustors. The effect of carryover varies based on the concentration of the high HCG sample. Recall start date: Mar 1, 2022

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Recall Numbers

64028

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected Products - Immulite 2000 System-Human Chorionic Gonadotropin (Hcg) Assay, Lot or Serial Number - 466, 463, Model or catalog number - L2KCG2, L2KCG6

Risk / hazard

Siemens Healthcare Diagnostics Inc. Has confirmed the potential for falsely elevated hcg results due to sample carryover. This can be observed when a sample is assayed for HCG immediately after an undiluted sample with a HCG value of GT;5000 MIU/mL. This issue impacts serum and urine patient samples, as well as quality control samples and adjustors. The effect of carryover varies based on the concentration of the high HCG sample. Recall start date: Mar 1, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Siemens, Immulite 2000 System-human Chorionic Gonadotropin Assay

Manufacturer

Not provided

Phone

Not provided

Brand: Siemens, Immulite 2000 System-human Chorionic Gonadotropin Assay.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

64028

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Rapidlab® 348ex system

Medium riskHealth Canada2 Jul 2026
CA
Recall image placeholder

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Medium riskHealth Canada18 Jun 2026
CA
Recall image placeholder

Dimension® creatinine (cre2)

Medium riskHealth Canada2 Apr 2026
CA
Recall image placeholder

Rapidpoint® 500 blood gas systems

Medium riskHealth Canada17 Mar 2026
CA
Recall image placeholder

Wash waste cartridge for rapidpoint® 500 and rapidpoint® 500e

Medium riskHealth Canada16 Jul 2025
CA
Recall image placeholder

Innovance® vwf ac assay

Medium riskHealth Canada4 Jul 2025
CA
Recall image placeholder

Atellica ch diluent

Medium riskHealth Canada26 Feb 2025
CA
Recall image placeholder

Atellica ci system

Medium riskHealth Canada15 Nov 2024
CA
Recall image placeholder

Dimension vista® system - total bilirubin assay

Medium riskHealth Canada21 Aug 2024
CA
Recall image placeholder

Rapidpoint® 500e blood gas system

Medium riskHealth Canada12 Oct 2023
CA
Recall image placeholder

Immulite 2000 sytstem thyroglobulin assay

Medium riskHealth Canada5 Jan 2023
CA
Recall image placeholder

Dimension exl system - cardiac troponin-i (ctni) assay

Medium riskHealth Canada13 Jun 2022

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 28 Mar 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.