No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
No image published for this recall
Dimension exl system - cardiac troponin-i (ctni) assay
Official notice
13 Jun 2022
An official Health Canada notice concerns Dimension exl system - cardiac troponin-i (ctni) assay. Siemens Healthcare Diagnostics Inc. Has confirmed that Dimension Cardiac Troponin I (lTni and CTnI) lots (ea2314 and ga2323) shows a positive bias with lithium heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/ml. Troponin values as high as 0.20 ng/ml have been observed with lithium heparin plasma that are expected to be within the 99th percentile. Serum samples collected from normal, healthy individuals whose results fall within the reference interval or below the 99th percentile of 0.07 ng/ml were not impacted by this issue. Recall start date: May 24, 2022 Published 2022-06-13.
What to do now
Contact the manufacturer if you require additional information.
Recall number
64256
Official notice
2022-06-13
The hazard
The hazard
Siemens Healthcare Diagnostics Inc. Has confirmed that Dimension Cardiac Troponin I (lTni and CTnI) lots (ea2314 and ga2323) shows a positive bias with lithium heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/ml. Troponin values as high as 0.20 ng/ml have been observed with lithium heparin plasma that are expected to be within the 99th percentile. Serum samples collected from normal, healthy individuals whose results fall within the reference interval or below the 99th percentile of 0.07 ng/ml were not impacted by this issue. Recall start date: May 24, 2022
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Siemens, Dimension Exl System - Cardiac Troponin-i Assay.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Atellica ch enzymatic hemoglobin a1c (a1c e) assay
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Rapidlab® 348ex system
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Dimension® creatinine (cre2)
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Rapidpoint® 500 blood gas systems
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Wash waste cartridge for rapidpoint® 500 and rapidpoint® 500e
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Innovance® vwf ac assay
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Atellica ch diluent
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Atellica ci system
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Dimension vista® system - total bilirubin assay
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Rapidpoint® 500e blood gas system
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Immulite 2000 sytstem thyroglobulin assay
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Immulite 2000 system-human chorionic gonadotropin (hcg) assay
Brand match
Medium risk
Markets
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 13 Jun 2022. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.