Open official notice

Gentle Threads, Interference Screw; Model No. 905617, for soft tissue reattachment procedures in the ankle foot and knee

Official notice

16 May 2018

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Gentle Threads, Interference Screw; Model No. 905617, for soft tissue reattachment procedures in the ankle foot and knee. Specific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle. Published 2018-05-16.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

54

Recall number

Z-1734-2018

Official notice

2018-05-16

The hazard

Specific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

207910

Event Ids

79682

Recall Numbers

Z-1734-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Gentle Threads, Interference Screw; Model No. 905617, for soft tissue reattachment procedures in the ankle/foot and knee.

Units affected

54

Hazard

Specific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

BR · Brazil

Affected area 4

CA · Canada

Affected area 5

CL · Chile

Affected area 6

CR · Costa Rica

Affected area 7

KR · South Korea

Affected area 8

MX · Mexico

Affected area 9

MY · Malaysia

Affected area 10

NL · Netherlands

Company & contacts

Manufacturer

Zimmer Biomet, Inc.

Phone

Not provided

Brand: Zimmer Biomet. Contact: Zimmer Biomet, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1734-2018

Event ID

79682

Units affected

54

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 16 May 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button