Filmarray 2.0 instrument, ref: flm2-asy-0001, flm2-asy-0001-w
Official notice
18 Sept 2024
An official FDA notice concerns Filmarray 2.0 instrument, ref: flm2-asy-0001, flm2-asy-0001-w. Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA. Published 2024-09-18.
What to do now
Do not consume or use the affected product. Return it to the place of purchase or follow instructions in the official FDA notice.
Units affected
1,001
Recall number
Z-3122-2024
Official notice
2024-09-18
The hazard
The hazard
Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Affected area 3
Affected area 4
Affected area 5
Affected area 6
Affected area 7
Affected area 8
Affected area 9
Affected area 10
Company & contacts
Company & contacts
Brand: Biofire Diagnostics. Contact: BioFire Diagnostics, LLC.
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.