Epoc blood analysis system - single use blood gas, electrolyte and metabolite test card
Official notice
15 Jun 2023
An official Health Canada notice concerns Epoc blood analysis system - single use blood gas, electrolyte and metabolite test card. Siemens Healthcare Diagnostics Inc. Has confirmed the occurrence of discrepant high glucose results in samples with glucose results on the lower end of the reportable range (at or below 54 mg/dl) when using the above listed lots of BGEM test cards in conjunction with sensor configuration 41.1 (software version: EPOC Host v3.37.3 and EPOC NXS v4.10.6). The observed maximum bias for the affected lots are detailed in the urgent field safety notice. Recall start date: June 5, 2023 Published 2023-06-15.
What to do now
Contact the manufacturer if you require additional information.
Recall number
73827
Official notice
2023-06-15
The hazard
The hazard
Siemens Healthcare Diagnostics Inc. Has confirmed the occurrence of discrepant high glucose results in samples with glucose results on the lower end of the reportable range (at or below 54 mg/dl) when using the above listed lots of BGEM test cards in conjunction with sensor configuration 41.1 (software version: EPOC Host v3.37.3 and EPOC NXS v4.10.6). The observed maximum bias for the affected lots are detailed in the urgent field safety notice. Recall start date: June 5, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Epocal, Epoc Blood Analysis System - Single Use Blood Gas Electrolyte And Metabolite Test Card.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Epoc blood analysis bgem test cards
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Epoc® blood analysis bgem test cards
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Epoc blood analysis bgem test card
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Epoc bgem bun test card [25pk]
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Epoc blood analysis bgem test cards
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Rapidlab® 348ex system
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Atellica ch enzymatic hemoglobin a1c (a1c e) assay
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Aptima hiv quant dx assay, hcv quant dx assay, and hbv quant assay software
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Vectorflow antegrade catheterization product
Brand match
Medium risk
Markets
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 15 Jun 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.