Epoc blood analysis system - single use blood gas, electrolyte and metabolite test card

Official notice

15 Jun 2023

Last checked

15 Jun 2023

Medium riskHealth Canada
Watch Epocal

An official Health Canada notice concerns Epoc blood analysis system - single use blood gas, electrolyte and metabolite test card. Siemens Healthcare Diagnostics Inc. Has confirmed the occurrence of discrepant high glucose results in samples with glucose results on the lower end of the reportable range (at or below 54 mg/dl) when using the above listed lots of BGEM test cards in conjunction with sensor configuration 41.1 (software version: EPOC Host v3.37.3 and EPOC NXS v4.10.6). The observed maximum bias for the affected lots are detailed in the urgent field safety notice. Recall start date: June 5, 2023 Published 2023-06-15.

What to do now

Contact the manufacturer if you require additional information.

Recall number

73827

Official notice

2023-06-15

The hazard

Siemens Healthcare Diagnostics Inc. Has confirmed the occurrence of discrepant high glucose results in samples with glucose results on the lower end of the reportable range (at or below 54 mg/dl) when using the above listed lots of BGEM test cards in conjunction with sensor configuration 41.1 (software version: EPOC Host v3.37.3 and EPOC NXS v4.10.6). The observed maximum bias for the affected lots are detailed in the urgent field safety notice. Recall start date: June 5, 2023

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Recall Numbers

73827

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - EPOC Blood Analysis BGEM Test Cards, Lot or serial number - 04-23059-50, Model or catalog number - 10736515

Risk / hazard

Siemens Healthcare Diagnostics Inc. Has confirmed the occurrence of discrepant high glucose results in samples with glucose results on the lower end of the reportable range (at or below 54 mg/dl) when using the above listed lots of BGEM test cards in conjunction with sensor configuration 41.1 (software version: EPOC Host v3.37.3 and EPOC NXS v4.10.6). The observed maximum bias for the affected lots are detailed in the urgent field safety notice. Recall start date: June 5, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Epocal, Epoc Blood Analysis System - Single Use Blood Gas Electrolyte And Metabolite Test Card

Manufacturer

Not provided

Phone

Not provided

Brand: Epocal, Epoc Blood Analysis System - Single Use Blood Gas Electrolyte And Metabolite Test Card.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

73827

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Epoc blood analysis bgem test cards

Medium riskHealth Canada13 May 2026
CA
Recall image placeholder

Epoc® blood analysis bgem test cards

Medium riskHealth Canada18 Jul 2025
CA
Recall image placeholder

Epoc blood analysis bgem test card

Medium riskHealth Canada9 Apr 2025
CA
Recall image placeholder

Epoc bgem bun test card [25pk]

Medium riskHealth Canada4 Oct 2024
CA
Recall image placeholder

Epoc blood analysis bgem test cards

Medium riskHealth Canada20 Jun 2022
CA
Recall image placeholder

Rapidlab® 348ex system

Medium riskHealth Canada2 Jul 2026
CA
Recall image placeholder

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Medium riskHealth Canada18 Jun 2026
CA
Recall image placeholder

Aptima hiv quant dx assay, hcv quant dx assay, and hbv quant assay software

Medium riskHealth Canada11 Jun 2026
CA
Recall image placeholder

Vectorflow antegrade catheterization product

Medium riskHealth Canada23 Apr 2026
CA
Recall image placeholder

I‑stat eg7+ cartridge

Critical riskHealth Canada15 Apr 2026
CA
Recall image placeholder

Quickvue® dipstick strep a test

Medium riskHealth Canada10 Apr 2026
CA
Recall image placeholder

Dimension® creatinine (cre2)

Medium riskHealth Canada2 Apr 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 15 Jun 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.