Epoc blood analysis bgem test cards

Official notice

20 Jun 2022

Last checked

20 Jun 2022

Medium riskHealth Canada
Watch Epocal

An official Health Canada notice concerns Epoc blood analysis bgem test cards. Siemens Healthcare Diagnostics Inc. Has confirmed the occurrence of inconsistent discrepant (low bias) po2 results. At a medical decision level of approximately 80mmhg, the maximum negative bias is - 11.7% (the acceptable bias is +/-10% for po2 range 50-250 mmhg). Recall start date: June 8, 2022 Published 2022-06-20.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64288

Official notice

2022-06-20

The hazard

Siemens Healthcare Diagnostics Inc. Has confirmed the occurrence of inconsistent discrepant (low bias) po2 results. At a medical decision level of approximately 80mmhg, the maximum negative bias is - 11.7% (the acceptable bias is +/-10% for po2 range 50-250 mmhg). Recall start date: June 8, 2022

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Recall Numbers

64288

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - Epoc Blood Analysis Bgem Test Cards, Lot or serial number - 112200550, Model or catalog number - 10736515

Risk / hazard

Siemens Healthcare Diagnostics Inc. Has confirmed the occurrence of inconsistent discrepant (low bias) po2 results. At a medical decision level of approximately 80mmhg, the maximum negative bias is - 11.7% (the acceptable bias is +/-10% for po2 range 50-250 mmhg). Recall start date: June 8, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Epocal, Epoc Blood Analysis Bgem Test Cards

Manufacturer

Not provided

Phone

Not provided

Brand: Epocal, Epoc Blood Analysis Bgem Test Cards.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

64288

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Epoc blood analysis bgem test cards

Medium riskHealth Canada13 May 2026
CA
Recall image placeholder

Epoc® blood analysis bgem test cards

Medium riskHealth Canada18 Jul 2025
CA
Recall image placeholder

Epoc blood analysis bgem test card

Medium riskHealth Canada9 Apr 2025
CA
Recall image placeholder

Epoc bgem bun test card [25pk]

Medium riskHealth Canada4 Oct 2024
CA
Recall image placeholder

Epoc blood analysis system - single use blood gas, electrolyte and metabolite test card

Medium riskHealth Canada15 Jun 2023
CA
Recall image placeholder

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Medium riskHealth Canada18 Jun 2026
CA
Recall image placeholder

Quickvue® dipstick strep a test

Medium riskHealth Canada10 Apr 2026
CA
Recall image placeholder

Rapidpoint® 500 blood gas systems

Medium riskHealth Canada17 Mar 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 20 Jun 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.