Epoc® blood analysis bgem test cards

Official notice

18 Jul 2025

Last checked

18 Jul 2025

Medium riskHealth Canada
Watch Epocal

An official Health Canada notice concerns Epoc® blood analysis bgem test cards. Siemens Healthcare Diagnostics Inc. Has confirmed that a negative sodium bias may exist in some epoc® test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/l. The maximum bias observed was -14 mmol/l, which occurred at a higher sodium concentration around 150 mmol/l. No positive bias has been observed. Quality control (qc) fluid may or may not detect the issue depending on the bias magnitude and the frequency of qc testing. Recall start date: July 9, 2025 Published 2025-07-18.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77740

Official notice

2025-07-18

The hazard

Siemens Healthcare Diagnostics Inc. Has confirmed that a negative sodium bias may exist in some epoc® test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/l. The maximum bias observed was -14 mmol/l, which occurred at a higher sodium concentration around 150 mmol/l. No positive bias has been observed. Quality control (qc) fluid may or may not detect the issue depending on the bias magnitude and the frequency of qc testing. Recall start date: July 9, 2025

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Recall Numbers

77740

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - epoc® Blood Analysis BGEM Test Cards, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - 10736515

Risk / hazard

Siemens Healthcare Diagnostics Inc. Has confirmed that a negative sodium bias may exist in some epoc® test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/l. The maximum bias observed was -14 mmol/l, which occurred at a higher sodium concentration around 150 mmol/l. No positive bias has been observed. Quality control (qc) fluid may or may not detect the issue depending on the bias magnitude and the frequency of qc testing. Recall start date: July 9, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Epocal, Epoc® Blood Analysis Bgem Test Cards

Manufacturer

Not provided

Phone

Not provided

Brand: Epocal, Epoc® Blood Analysis Bgem Test Cards.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

77740

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Epoc blood analysis bgem test cards

Medium riskHealth Canada13 May 2026
CA
Recall image placeholder

Epoc blood analysis bgem test card

Medium riskHealth Canada9 Apr 2025
CA
Recall image placeholder

Epoc bgem bun test card [25pk]

Medium riskHealth Canada4 Oct 2024
CA
Recall image placeholder

Epoc blood analysis system - single use blood gas, electrolyte and metabolite test card

Medium riskHealth Canada15 Jun 2023
CA
Recall image placeholder

Epoc blood analysis bgem test cards

Medium riskHealth Canada20 Jun 2022
CA
Recall image placeholder

Rapidlab® 348ex system

Medium riskHealth Canada2 Jul 2026
CA
Recall image placeholder

Wasplab

Medium riskHealth Canada2 Jul 2026
CA
Recall image placeholder

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Medium riskHealth Canada18 Jun 2026
CA
Recall image placeholder

Cannapiece corp. Recalls wild west 96+ strawberry stampede liquid diamonds 510 vape due to a labelling issue

Critical riskHealth Canada10 Jun 2026
CA
Recall image placeholder

Red rubber latex catheters

Medium riskHealth Canada22 May 2026
CA
Recall image placeholder

Neurological sponges

Medium riskHealth Canada21 May 2026
CA
Recall image placeholder

Cannapiece corp. Recalls jays - pink pennifer 0.4g all-in-one vape due to labelling issue

Critical riskHealth Canada20 May 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 18 Jul 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.