No image published for this recall

No image published for this recall

Easy moving plus b mobile x-ray unit with digital detector

Official notice

20 Jun 2024

Last checked

20 Jun 2024

Medium riskHealth Canada

An official Health Canada notice concerns Easy moving plus b mobile x-ray unit with digital detector. Notification to inform customers regarding potential safety issue related to the steel cable installed inside the mobile column which supports the weight of telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm of the cable is cut off does not work properly. The interference occurs between the cable and a metal plate of the arm's carriage due to shape of the cable's crimped terminals. Scope of the safety recall is to review the installed base according to the manufacturer's instructions and install the kit immediately if the cable could interfere with the parachute system. Recall start date: June 5, 2024 Published 2024-06-20.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75707

Official notice

2024-06-20

The hazard

Notification to inform customers regarding potential safety issue related to the steel cable installed inside the mobile column which supports the weight of telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm of the cable is cut off does not work properly. The interference occurs between the cable and a metal plate of the arm's carriage due to shape of the cable's crimped terminals. Scope of the safety recall is to review the installed base according to the manufacturer's instructions and install the kit immediately if the cable could interfere with the parachute system. Recall start date: June 5, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75707

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Easy Moving Plus B Mobile X-Ray Unit With Digital Detector, Lot or serial number - G71467 G77408 G73624 G82557 G79899 G82499, Model or catalogue number - SM-40HF-B-D-VIR

Risk / hazard

Notification to inform customers regarding potential safety issue related to the steel cable installed inside the mobile column which supports the weight of telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm of the cable is cut off does not work properly. The interference occurs between the cable and a metal plate of the arm's carriage due to shape of the cable's crimped terminals. Scope of the safety recall is to review the installed base according to the manufacturer's instructions and install the kit immediately if the cable could interfere with the parachute system. Recall start date: June 5, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Sociedad Espanola De Electromedicina Y Calidad Technidata S.a, Easy Moving Plus B Mobile X-ray Unit With Digital Detector.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

75707

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Abiomed introducer kits with impella heart pumps

High riskHealth Canada29 Jun 2026
CA
Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)Image: Health Canada recall notice

Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Medium riskHealth Canada22 Jun 2026
CA
No image published for this recall

Penumbra system red 68 reperfusion catheter

Critical riskHealth Canada22 Jun 2026
CA
No image published for this recall

Faradrive™ steerable sheath

Medium riskHealth Canada19 Jun 2026
CA
No image published for this recall

Ge healthcare integrated and stand-alone resuscitation systems

Medium riskHealth Canada16 Jun 2026
CA
No image published for this recall

Unauthorized soft-shelled hyperbaric chambers may pose serious health risks 2026-06-16

Critical riskHealth Canada16 Jun 2026
CA
No image published for this recall

Etms 70 mm cooled coil

Medium riskHealth Canada11 Jun 2026
CA
No image published for this recall

Omnipod® insulin management system - omnipod®

Critical riskHealth Canada1 Jun 2026
CA
No image published for this recall

Relay®pro thoracic stent-graft system

Critical riskHealth Canada29 May 2026
CA
No image published for this recall

Merlin™ 3650 patient care system (pcs)

Medium riskHealth Canada22 May 2026
CA
No image published for this recall

Hard disk drives in azurion and allura systems

Critical riskHealth Canada15 May 2026
CA
No image published for this recall

Impella controller

Critical riskHealth Canada13 May 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 20 Jun 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.