Open official notice

Compress, Mini Compress; Item Nos. 178350 178351 178352 178353 178354 178355 178356 178357 178358 178359 178360 178361 178362 178363 178364

Official notice

4 Dec 2019

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Compress, Mini Compress; Item Nos. 178350 178351 178352 178353 178354 178355 178356 178357 178358 178359 178360 178361 178362 178363 178364. Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. Published 2019-12-04.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

219,988

Recall number

Z-0519-2020

Official notice

2019-12-04

The hazard

Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

83783

Recall Numbers

Z-0519-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Compress, Mini Compress; Item Nos. 178350 178351 178352 178353 178354 178355 178356 178357 178358 178359 178360 178361 178362 178363 178364 178365 178366 178367 178368 178369 178370 178371 178372 178373 178464 178472 178480 178488 178496 178504 178575 178576 178577 178578 178579 178580 178730 178731 178732 178733 178734 178735 178354S 178754 178755 178756 178757 178758 178759 Product Usage: 1) Correction of revision of unsuccessful osteotomy, arthrodesis or previous joint replacement 2) Tumor resections 3) Revision of previously failed total joint arthroplasty 4) Trauma The ComPreSs Distal Femoral Replacement System is intended for uncemented use.

Units affected

219,988

Hazard

Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AU · Australia

Affected area 4

BR · Brazil

Affected area 5

CA · Canada

Affected area 6

CL · Chile

Affected area 7

CO · Colombia

Affected area 8

CR · Costa Rica

Affected area 9

IN · India

Affected area 10

IT · Italy

Affected area 11

JP · Japan

Affected area 12

KR · South Korea

Affected area 13

MX · Mexico

Affected area 14

MY · Malaysia

Affected area 15

NL · Netherlands

Affected area 16

NZ · New Zealand

Affected area 17

SG · Singapore

Affected area 18

SV · El Salvador

Affected area 19

TH · Thailand

Affected area 20

TW · Taiwan

Affected area 21

VE · Venezuela

Company & contacts

Manufacturer

Zimmer Biomet, Inc.

Phone

Not provided

Brand: Zimmer Biomet. Contact: Zimmer Biomet, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0519-2020

Event ID

83783

Units affected

219,988

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 4 Dec 2019. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button