Open official notice

CIVCO lnfiniti Needle Guide Bracket. Used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound

Official notice

29 Aug 2018

RecallScope risk estimate:MediumFDA

CIVCO lnfiniti Needle Guide Bracket. Used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers. Product Usage: The bracket and disposable guide provide physicians with a tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

177

Recall number

Z-2888-2018

Official notice

2018-08-29

The hazard

A cover failed leak testing during validation testing of new needle guides from a replacement tool.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

80704

Recall Numbers

Z-2888-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

CIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers. Product Usage: The bracket and disposable guide provide physicians with a tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers.

Units affected

177

Hazard

A cover failed leak testing during validation testing of new needle guides from a replacement tool.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

BE · Belgium

Affected area 3

CA · Canada

Affected area 4

CH · Switzerland

Affected area 5

CN · China

Affected area 6

DE · Germany

Affected area 7

DK · Denmark

Affected area 8

ES · Spain

Affected area 9

GB · United Kingdom

Affected area 10

HK · Hong Kong

Affected area 11

IT · Italy

Affected area 12

JP · Japan

Affected area 13

NL · Netherlands

Affected area 14

UA · Ukraine

Company & contacts

Manufacturer

Civco Medical Instruments Co. Inc.

Phone

Not provided

Brand: Civco Medical Instruments. Contact: Civco Medical Instruments Co. Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2888-2018

Event ID

80704

Units affected

177

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Aug 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button