No image published for this recall

Centrimag™ adult ecmo kit and blood pump

Official notice

22 Oct 2025

CriticalHealth Canada

An official Health Canada notice concerns Centrimag™ adult ecmo kit and blood pump. Abbott is initiating a voluntary recall customer notification to remind customers that when using the CentriMag™ Acute Circulatory Support System you must always ensure the CentriMag™ blood pump is securely locked into the CentriMag™ motor as described in the instructions for use (IFU). Failure to properly lock the blood pump into the motor could lead to particulate or increased heat while using the system. At this time, Abbott is not requesting that this product be returned or removed from the field. If the pump is able to be locked into the motor, the product is safe for use and functions as intended when the IFU is followed to securely seat the pump. Published 2025-10-22.

What to do now

Contact the manufacturer if you require additional information.

Recall number

78349

Official notice

2025-10-22

The hazard

Abbott is initiating a voluntary recall customer notification to remind customers that when using the CentriMag™ Acute Circulatory Support System you must always ensure the CentriMag™ blood pump is securely locked into the CentriMag™ motor as described in the instructions for use (IFU). Failure to properly lock the blood pump into the motor could lead to particulate or increased heat while using the system. At this time, Abbott is not requesting that this product be returned or removed from the field. If the pump is able to be locked into the motor, the product is safe for use and functions as intended when the IFU is followed to securely seat the pump.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

78349

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - CentriMag™ Blood Pump, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 201-90016

Product 2

Affected products - CentriMag™ Adult ECMO Kit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - CMAEK01

Risk / hazard

Abbott is initiating a voluntary recall customer notification to remind customers that when using the CentriMag™ Acute Circulatory Support System you must always ensure the CentriMag™ blood pump is securely locked into the CentriMag™ motor as described in the instructions for use (IFU). Failure to properly lock the blood pump into the motor could lead to particulate or increased heat while using the system. At this time, Abbott is not requesting that this product be returned or removed from the field. If the pump is able to be locked into the motor, the product is safe for use and functions as intended when the IFU is followed to securely seat the pump.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Abbott Medical, Centrimag™ Adult Ecmo Kit And Blood Pump.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

78349

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

HEALTH
Medium

Abbotts proclaim™ scs proclaim™ drg infinity™ dbs implantable pulse generators

Health Canada30 Apr 2026

Brand match

Medium risk

Markets

CA
HEALTH
High

Heartmate 3™ system controller

Health Canada17 Oct 2025

Brand match

High risk

Markets

CA
HEALTH
Medium

Mobile power unit

Health Canada9 Jul 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Eterna™ implantable pulse generator

Health Canada15 Oct 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Tactiflex™ ablation catheter, sensor enabled™

Health Canada17 Sept 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Amplatzer torqvue delivery system

Health Canada26 Mar 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Heartmate 3 lvas implant kit and heartmate 3 system controller

Health Canada30 Jul 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

2Nd generation centrimag™ primary console with em-tec adult flow probe

Health Canada14 Jun 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Proclaim and infinity implantable pulse generators

Health Canada15 Aug 2023

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Cardiomems patient and hospital electronics system

Health Canada23 Feb 2023

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Cardiomems hospital electronics system

Health Canada23 Feb 2023

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Tacticath contact force ablation catheter, sensor enabled

Health Canada4 Jul 2022

Brand match

Medium risk

Markets

CA

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 22 Oct 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.