No image published for this recall

Proclaim and infinity implantable pulse generators

Official notice

15 Aug 2023

MediumHealth Canada

An official Health Canada notice concerns Proclaim and infinity implantable pulse generators. Abbott has received complaints from patients who are unable to exit MRI mode, as their patient controller has lost the ability to connect or communicate with their IPG while in MRI mode. Abbott sent a customer communication on 2023-07-21 to remind health care providers about use of the magnetic resonance imaging (MRI) mode feature on these systems and the associated risks related to inability to exit MRI mode, including the potential need for additional surgery to restore therapy. Recall start date: July 21, 2023 Published 2023-08-15.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74120

Official notice

2023-08-15

The hazard

Abbott has received complaints from patients who are unable to exit MRI mode, as their patient controller has lost the ability to connect or communicate with their IPG while in MRI mode. Abbott sent a customer communication on 2023-07-21 to remind health care providers about use of the magnetic resonance imaging (MRI) mode feature on these systems and the associated risks related to inability to exit MRI mode, including the potential need for additional surgery to restore therapy. Recall start date: July 21, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

74120

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Proclaim Drg Implantable Pulse Generator, Lot or serial number - All lots., Model or catalogue number - 3664

Product 2

Affected products - Infinity 5 Implantable Pulse Generator, Lot or serial number - All lots., Model or catalogue number - 6660

Product 3

Affected products - Proclaim Xr 7 Implantable Pulse Generator, Lot or serial number - All lots., Model or catalogue number - 3662

Product 4

Affected products - Infinity 7 Implantable Pulse Generator, Lot or serial number - All lots., Model or catalogue number - 6662

Product 5

Affected products - Proclaim Xr 5 Implantable Pulse Generator, Lot or serial number - All lots., Model or catalogue number - 3660

Risk / hazard

Abbott has received complaints from patients who are unable to exit MRI mode, as their patient controller has lost the ability to connect or communicate with their IPG while in MRI mode. Abbott sent a customer communication on 2023-07-21 to remind health care providers about use of the magnetic resonance imaging (MRI) mode feature on these systems and the associated risks related to inability to exit MRI mode, including the potential need for additional surgery to restore therapy. Recall start date: July 21, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Abbott Medical, Proclaim And Infinity Implantable Pulse Generators.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

74120

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

HEALTH
Medium

Abbotts proclaim™ scs proclaim™ drg infinity™ dbs implantable pulse generators

Health Canada30 Apr 2026

Brand match

Medium risk

Markets

CA
HEALTH
Critical

Centrimag™ adult ecmo kit and blood pump

Health Canada22 Oct 2025

Brand match

Critical risk

Markets

CA
HEALTH
High

Heartmate 3™ system controller

Health Canada17 Oct 2025

Brand match

High risk

Markets

CA
HEALTH
Medium

Eterna™ implantable pulse generator

Health Canada15 Oct 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Tactiflex™ ablation catheter, sensor enabled™

Health Canada17 Sept 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Mobile power unit

Health Canada9 Jul 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Amplatzer torqvue delivery system

Health Canada26 Mar 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Cardiomems patient electronics system

Health Canada18 Oct 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Aveir™ leadless pacemaker

Health Canada19 Aug 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Heartmate 3 lvas implant kit and heartmate 3 system controller

Health Canada30 Jul 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Proclaim implantable pulse generators

Health Canada4 Jul 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Infinity 5 and 7 implantable pulse generator

Health Canada3 Jul 2024

Brand match

Medium risk

Markets

CA

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 15 Aug 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.