Cardiomems hospital electronics system

Official notice

23 Feb 2023

Last checked

23 Feb 2023

Medium riskHealth Canada
Watch Abbott Medical

An official Health Canada notice concerns Cardiomems hospital electronics system. The Cardiomems HF System is intended to operate at frequencies between 30 and 37.5 MHz. Abbott has found that a small number (less than 1%) of implanted Cardiomems PA sensors (model CM2000) have operated outside of this intended radiofrequency range at least once over the life of the implant when interrogated by Cardiomems patient or hospital systems (models CM1100 and CM3000). Operating frequency may be affected by atmospheric conditions, elevation, pulmonary artery (pa) pressure, and unique sensor and patient characteristics. Recall start date: February 7, 2023 Published 2023-02-23.

What to do now

Contact the manufacturer if you require additional information.

Recall number

72929

Official notice

2023-02-23

The hazard

The Cardiomems HF System is intended to operate at frequencies between 30 and 37.5 MHz. Abbott has found that a small number (less than 1%) of implanted Cardiomems PA sensors (model CM2000) have operated outside of this intended radiofrequency range at least once over the life of the implant when interrogated by Cardiomems patient or hospital systems (models CM1100 and CM3000). Operating frequency may be affected by atmospheric conditions, elevation, pulmonary artery (pa) pressure, and unique sensor and patient characteristics. Recall start date: February 7, 2023

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Recall Numbers

72929

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - Cardiomems Hospital Electronics System, Lot or serial number - All lots., Model or catalog number - CM2000

Product 2

Affected products - Cardiomems Hospital Electronics System, Lot or serial number - All lots., Model or catalog number - CM3000

Risk / hazard

The Cardiomems HF System is intended to operate at frequencies between 30 and 37.5 MHz. Abbott has found that a small number (less than 1%) of implanted Cardiomems PA sensors (model CM2000) have operated outside of this intended radiofrequency range at least once over the life of the implant when interrogated by Cardiomems patient or hospital systems (models CM1100 and CM3000). Operating frequency may be affected by atmospheric conditions, elevation, pulmonary artery (pa) pressure, and unique sensor and patient characteristics. Recall start date: February 7, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Abbott Medical, Cardiomems Hospital Electronics System

Manufacturer

Not provided

Phone

Not provided

Brand: Abbott Medical, Cardiomems Hospital Electronics System.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

72929

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Abbotts proclaim™ scs proclaim™ drg infinity™ dbs implantable pulse generators

Medium riskHealth Canada30 Apr 2026
CA
Recall image placeholder

Centrimag™ adult ecmo kit and blood pump

Critical riskHealth Canada22 Oct 2025
CA
Recall image placeholder

Heartmate 3™ system controller

High riskHealth Canada17 Oct 2025
CA
Recall image placeholder

Eterna™ implantable pulse generator

Medium riskHealth Canada15 Oct 2025
CA
Recall image placeholder

Tactiflex™ ablation catheter, sensor enabled™

Medium riskHealth Canada17 Sept 2025
CA
Recall image placeholder

Mobile power unit

Medium riskHealth Canada9 Jul 2025
CA
Recall image placeholder

Amplatzer torqvue delivery system

Medium riskHealth Canada26 Mar 2025
CA
Recall image placeholder

Cardiomems patient electronics system

Medium riskHealth Canada18 Oct 2024
CA
Recall image placeholder

Aveir™ leadless pacemaker

Medium riskHealth Canada19 Aug 2024
CA
Recall image placeholder

Heartmate 3 lvas implant kit and heartmate 3 system controller

Medium riskHealth Canada30 Jul 2024
CA
Recall image placeholder

Proclaim implantable pulse generators

Medium riskHealth Canada4 Jul 2024
CA
Recall image placeholder

Infinity 5 and 7 implantable pulse generator

Medium riskHealth Canada3 Jul 2024

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 23 Feb 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.