No image published for this recall

No image published for this recall

Centricity™ high acuity anesthesia (cha-a)

Official notice

17 Jun 2024

Last checked

17 Jun 2024

Medium riskHealth Canada

An official Health Canada notice concerns Centricity™ high acuity anesthesia (cha-a). Certain drug and fluid administration details may be missing from trend view in the Centricity™ High Acuity Anesthesia (CHA-A) system. Recall start date: June 7, 2024 Published 2024-06-17.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75688

Official notice

2024-06-17

The hazard

Certain drug and fluid administration details may be missing from trend view in the Centricity™ High Acuity Anesthesia (CHA-A) system. Recall start date: June 7, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75688

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Centricity High Acuity Anesthesia (CHA-A), Lot or serial number - 106494 JGHCHAA 106494TST 105084, Model or catalog number - 5854438-004 K1900EJED

Risk / hazard

Certain drug and fluid administration details may be missing from trend view in the Centricity™ High Acuity Anesthesia (CHA-A) system. Recall start date: June 7, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: GE Healthcare Iits Also Doing Business GE Healthcare, Centricity™ High Acuity Anesthesia.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

75688

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 17 Jun 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.