Open official notice

Carescape central station

Official notice

26 Aug 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Carescape central station. GE Healthcare has become aware that the CARESCAPE Central Station (CSCS) v3.0.x may not resume patient monitoring for at least 3 minutes following a system restart. Published 2026-08-26.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82531

Official notice

2026-08-26

The hazard

GE Healthcare has become aware that the CARESCAPE Central Station (CSCS) v3.0.x may not resume patient monitoring for at least 3 minutes following a system restart.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

5867474-04, 5867474-03

Other

RA RA-82531

Model numbers

5867474-04, 5867474-03

Recall Numbers

82531

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 5867474-04

Product 2

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 5867474-03

Hazard

GE Healthcare has become aware that the CARESCAPE Central Station (CSCS) v3.0.x may not resume patient monitoring for at least 3 minutes following a system restart.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: GE Medical Information.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82531

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 26 Aug 2026. The next one won’t wait for you to search.