Open official notice

Biofire® filmarray® 2.0 instruments and biofire® filmarray® torch module box

Official notice

29 Aug 2024

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Biofire® filmarray® 2.0 instruments and biofire® filmarray® torch module box. Biomérieux has identified an issue with a subset of internal equipment used during the temperature calibration process on BioFire torch modules and 2.0 instruments manufactured or serviced since May 2022. The investigation found that BioFire instruments calibrated with this equipment may report higher melting temperatures (TMS). The impacted instruments must be returned to a BioFire service center for re-calibration. Recall Start Date: August 9, 2024 Published 2024-08-29.

What to do now

Contact the importer, bioMérieux Canada.

Recall number

75985

Official notice

2024-08-29

The hazard

Biomérieux has identified an issue with a subset of internal equipment used during the temperature calibration process on BioFire torch modules and 2.0 instruments manufactured or serviced since May 2022. The investigation found that BioFire instruments calibrated with this equipment may report higher melting temperatures (TMS). The impacted instruments must be returned to a BioFire service center for re-calibration. Recall Start Date: August 9, 2024

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

75985

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - BioFire® FilmArray® 2.0 Instruments, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - FLM2-ASY-0001

Product 2

Affected products - BioFire® FilmArray® Torch Module Box, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - HTFA-SUB-0103

Hazard

Biomérieux has identified an issue with a subset of internal equipment used during the temperature calibration process on BioFire torch modules and 2.0 instruments manufactured or serviced since May 2022. The investigation found that BioFire instruments calibrated with this equipment may report higher melting temperatures (TMS). The impacted instruments must be returned to a BioFire service center for re-calibration. Recall Start Date: August 9, 2024

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Biofire Diagnostics, Biofire® Filmarray® 2.0 Instruments And Biofire® Filmarray® Torch Module Box.

Source & status

Tier

official

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

75985

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Aug 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button