No image published for this recall
Bd veritor plus analyzer
Official notice
17 Jan 2022
Last checked
17 Jan 2022
An official Health Canada notice concerns Bd veritor plus analyzer. Bd has received two (2) complaints reporting that the Bd Veritor Plus analyzer may have the potential to overheat and/or cause fire. Bd's investigation has found that the issue can occur when an unauthorized alternative ac power adapter, and not the one provided by Bd with the Bd Veritor Plus analyzer, is used to power the instrument. Recall start date: Jan 5, 2022 Published 2022-01-17.
What to do now
Contact the manufacturer if you require additional information.
Recall number
63828
Official notice
2022-01-17
The hazard
The hazard
Bd has received two (2) complaints reporting that the Bd Veritor Plus analyzer may have the potential to overheat and/or cause fire. Bd's investigation has found that the issue can occur when an unauthorized alternative ac power adapter, and not the one provided by Bd with the Bd Veritor Plus analyzer, is used to power the instrument. Recall start date: Jan 5, 2022
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Becton Dickinson And [bd], Bd Veritor Plus Analyzer.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Bd synapsys™ informatics solution

Bd phoenix™

Bd synapsys

Bd synapsys

Bd synapsys

Intramedullary bone saw blade assembly

Glidesheath slender® introducer sheath
Image: Health Canada recall noticeImportation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Unauthorized soft-shelled hyperbaric chambers may pose serious health risks 2026-06-16

Harmony™ delivery catheter system

Relay®pro thoracic stent-graft system

Lifepak 15 v4 spco spmet sensors
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 17 Jan 2022. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.