Open official notice

Altrx Altralinked Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to

Official notice

29 Aug 2012

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Altrx Altralinked Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to. In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner. Published 2012-08-29.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

20

Recall number

Z-2220-2012

Official notice

2012-08-29

The hazard

In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

231859

Event Ids

62649

Recall Numbers

Z-2220-2012

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to replace the damaged hip joint to help increase patient mobility and to reduce pain. Packaging: Product is placed into a PETG tray, sealed with a Tyvek lid. This is placed into a second PETG tray and sealed with a Tyvek lid. This package is placed into a paperboard carton, labeled, and shrink-wrapped. Intended Use: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Units affected

20

Hazard

In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · PA

Affected area 1

US · United States

Affected area 2

CA · Canada

Company & contacts

Manufacturer

DePuy Orthopaedics, Inc.

Phone

Not provided

Brand: Depuy Orthopaedics. Contact: DePuy Orthopaedics, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2220-2012

Event ID

62649

Units affected

20

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Aug 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button