Alinity i hbsag reagent kit
Official notice
8 Jan 2024
Last checked
8 Jan 2024
An official Health Canada notice concerns Alinity i hbsag reagent kit. Abbott has identified a few cartridges within Alinity I HBsAg reagent kit, list number 08P0852, lot 51503FN00 that may exhibit variability in relative light unit (RLU) response and concentration values which may result in controls out of range, and/or incorrect patient results. Positive control out of range low results and/or negative control out of range high results could occur when using an impacted cartridge. Per the instructions for use (IFU), controls are required to be run once within each 24 hours of use. A potential falsely decreased or falsely increased patient result may occur for any of the following scenarios: 1. An impacted cartridge was used without running controls. 2. An impacted cartridge was used following calibration with a non-impacted cartridge. 3. An impacted cartridge was successfully calibrated, and controls were within range but calibrator and/or control RLUs are low. Recall start date: December 21, 2023 Published 2024-01-08.
What to do now
Contact the manufacturer if you require additional information.
Recall number
74882
Official notice
2024-01-08
The hazard
The hazard
Abbott has identified a few cartridges within Alinity I HBsAg reagent kit, list number 08P0852, lot 51503FN00 that may exhibit variability in relative light unit (RLU) response and concentration values which may result in controls out of range, and/or incorrect patient results. Positive control out of range low results and/or negative control out of range high results could occur when using an impacted cartridge. Per the instructions for use (IFU), controls are required to be run once within each 24 hours of use. A potential falsely decreased or falsely increased patient result may occur for any of the following scenarios: 1. An impacted cartridge was used without running controls. 2. An impacted cartridge was used following calibration with a non-impacted cartridge. 3. An impacted cartridge was successfully calibrated, and controls were within range but calibrator and/or control RLUs are low. Recall start date: December 21, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Lots
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Abbott Laboratories Diagnostics Division, Alinity I Hbsag Reagent Kit.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Architect i system - stat myoglobin assay

Solar™ gi

Rapidlab® 348ex system

Lifecodes lsa class ii

Nephromax high pressure nephrostomy balloon catheter

A.t.s® 4000 ts tourniquet systems single and dual hose with cpc connectors

Atellica ch enzymatic hemoglobin a1c (a1c e) assay
Image: Health Canada recall noticeCannapiece corp. Recalls wild west 96+ strawberry stampede liquid diamonds 510 vape due to a labelling issue

Iceman classic3™

Gsp® neonatal galt kit

Harmonyair a-series surgical lighting system and alyon surgical lighting system

Neurological sponges
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 8 Jan 2024. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.