Alaris & bd alaris point-of-care unit (pcu)

Official notice

20 Feb 2026

Last checked

20 Feb 2026

Critical riskHealth Canada
Watch Carefusion 303

An official Health Canada notice concerns Alaris & bd alaris point-of-care unit (pcu). Becton Dickinson (BD) is issuing this voluntary product communication because its labeling does not indicate which Transport Layer Security (TLS) versions must be enabled on the hospital network to ensure secure wi-fi communication with the Alaris and BD Alaris™ Point-of-Care Unit (PCU) when using enterprise-level wi-fi security. Without clear labeling, hospitals may unknowingly disable TLS 1.0, resulting in network connectivity failure and memory leaks which could lead to potential system error. Alaris™ and BD Alaris™ PCUs (v12.1.3 and prior) are not compatible with TLS security frameworks beyond TLS 1.0. If TLS 1.0 is disabled on the customer network, the PCUs will fail to connect to the wireless network. For customers utilizing bi-directional workflow, this leads to interoperability failures which require users to manually program devices. If TLS 1.0 is disabled on the customer network during use, PCUs (V12.1.3 and prior) may encounter a memory leak when the authentication fails, which consumes over time, all available memory and leads to a system error. This may happen during active infusions. Future versions of BD Alaris™ PCUs that are not yet available in Canada will be compatible with TLS 1.0, 1.1, and 1.2. Published 2026-02-20.

What to do now

Contact the manufacturer if you require additional information.

Recall number

81629

Official notice

2026-02-20

The hazard

Becton Dickinson (BD) is issuing this voluntary product communication because its labeling does not indicate which Transport Layer Security (TLS) versions must be enabled on the hospital network to ensure secure wi-fi communication with the Alaris and BD Alaris™ Point-of-Care Unit (PCU) when using enterprise-level wi-fi security. Without clear labeling, hospitals may unknowingly disable TLS 1.0, resulting in network connectivity failure and memory leaks which could lead to potential system error. Alaris™ and BD Alaris™ PCUs (v12.1.3 and prior) are not compatible with TLS security frameworks beyond TLS 1.0. If TLS 1.0 is disabled on the customer network, the PCUs will fail to connect to the wireless network. For customers utilizing bi-directional workflow, this leads to interoperability failures which require users to manually program devices. If TLS 1.0 is disabled on the customer network during use, PCUs (V12.1.3 and prior) may encounter a memory leak when the authentication fails, which consumes over time, all available memory and leads to a system error. This may happen during active infusions. Future versions of BD Alaris™ PCUs that are not yet available in Canada will be compatible with TLS 1.0, 1.1, and 1.2.

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Skus

8015

Other

RA RA-81629

Model Numbers

8015

Recall Numbers

81629

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 8015

Risk / hazard

Becton Dickinson (BD) is issuing this voluntary product communication because its labeling does not indicate which Transport Layer Security (TLS) versions must be enabled on the hospital network to ensure secure wi-fi communication with the Alaris and BD Alaris™ Point-of-Care Unit (PCU) when using enterprise-level wi-fi security. Without clear labeling, hospitals may unknowingly disable TLS 1.0, resulting in network connectivity failure and memory leaks which could lead to potential system error. Alaris™ and BD Alaris™ PCUs (v12.1.3 and prior) are not compatible with TLS security frameworks beyond TLS 1.0. If TLS 1.0 is disabled on the customer network, the PCUs will fail to connect to the wireless network. For customers utilizing bi-directional workflow, this leads to interoperability failures which require users to manually program devices. If TLS 1.0 is disabled on the customer network during use, PCUs (V12.1.3 and prior) may encounter a memory leak when the authentication fails, which consumes over time, all available memory and leads to a system error. This may happen during active infusions. Future versions of BD Alaris™ PCUs that are not yet available in Canada will be compatible with TLS 1.0, 1.1, and 1.2.

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Carefusion 303

Manufacturer

Not provided

Phone

Not provided

Brand: Carefusion 303.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

81629

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Bd alaris™ pump module

Medium riskHealth Canada29 Oct 2025
CA
Recall image placeholder

Bd alaris™ pump module and subset of compatible pump infusion sets

Medium riskHealth Canada24 Sept 2025
CA
Recall image placeholder

Alaris™ pump module & bd alaris™ pump module

Critical riskHealth Canada22 Jul 2025
CA
Recall image placeholder

Bd alaris pcu

Medium riskHealth Canada11 Sept 2024
CA
Recall image placeholder

Bd alaris™ etco2, pca, syringe, and pump modules

Medium riskHealth Canada30 Aug 2024
CA
Recall image placeholder

Bd alaris™ infusion devices

High riskHealth Canada30 Oct 2023
CA
Recall image placeholder

Bd alaris large volume pump

Critical riskHealth Canada1 Dec 2022
CA
Recall image placeholder

Bd alaris

Medium riskHealth Canada28 Feb 2022
CA
Recall image placeholder

Heartsine® samaritan® pad sam 350p sam 360p sam 500p - membrane tail

Medium riskHealth Canada23 Jul 2026
CA
Recall image placeholder

Hybrid hydrophilic microguidewire

Medium riskHealth Canada3 Jul 2026
CA
Recall image placeholder

Abiomed introducer kits with impella heart pumps

High riskHealth Canada29 Jun 2026
CA
Recall image placeholder

Infinion™ cx contact lead kit

Critical riskHealth Canada24 Jun 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 20 Feb 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.