Open official notice

Adaptor 3602, connector for hilger neurostim dual

Official notice

2 Oct 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Adaptor 3602, connector for hilger neurostim dual. Adapter #3602 is being recalled due to potential mechanical instability between adapter #3602 and probe #3602-00-TE when used with the Hilger Facial Nerve Stimulator, leading to physical detachment and potential loss of electrical contact during nerve monitoring. Published 2026-10-02.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82701

Official notice

2026-10-02

The hazard

Adapter #3602 is being recalled due to potential mechanical instability between adapter #3602 and probe #3602-00-TE when used with the Hilger Facial Nerve Stimulator, leading to physical detachment and potential loss of electrical contact during nerve monitoring.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

RA RA-82701

Recall Numbers

82701

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - ADAP.3602

Hazard

Adapter #3602 is being recalled due to potential mechanical instability between adapter #3602 and probe #3602-00-TE when used with the Hilger Facial Nerve Stimulator, leading to physical detachment and potential loss of electrical contact during nerve monitoring.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada · Quebec

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Distribution Instrumentarium Dba Instrumentarium.

Source & status

Tier

beta

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

82701

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 2 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button