Open official notice

Xcela picc kit with endexo and pasv valve technology, xcela pasv 5f dl 55cm mst-70 nitinol kit pg, upn h965251250, catalog no. 25-125 the

Official notice

20 Dec 2017

RecallScope risk estimate:MediumFDA

Xcela PICC Kit with ENDEXO and PASV Valve Technology, XCELA PASV 5F DL 55CM MST-70 Nitinol KIT PG, UPN H965251250, Catalog No. 25-125 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

4,591

Recall number

Z-0240-2018

Official notice

2017-12-20

The hazard

Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

4908974

Event Ids

76570

Recall Numbers

Z-0240-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Xcela PICC Kit with ENDEXO and PASV Valve Technology, XCELA PASV 5F DL 55CM MST-70 Nitinol KIT PG, UPN H965251250, Catalog No. 25-125 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.

Units affected

4,591

Hazard

Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Affected area 4

ES · Spain

Affected area 5

GB · United Kingdom

Affected area 6

NL · Netherlands

Company & contacts

Manufacturer

Angiodynamics Inc. (Navilyst Medical Inc.)

Phone

Not provided

Brand: Angiodynamics. Contact: Angiodynamics Inc. (Navilyst Medical Inc.).

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0240-2018

Event ID

76570

Units affected

4,591

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 20 Dec 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button