VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to calibrate
Official notice
19 Jul 2023
An official FDA notice concerns VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to calibrate. Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could be reported depending on the analyte and/or sample concentration. Published 2023-07-19.
What to do now
Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.
Units affected
2,400
Recall number
Z-2129-2023
Official notice
2023-07-19
The hazard
The hazard
Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could be reported depending on the analyte and/or sample concentration.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Other
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Affected area 3
Affected area 4
Affected area 5
Affected area 6
Affected area 7
Affected area 8
Affected area 9
Affected area 10
Affected area 11
Affected area 12
Affected area 13
Affected area 14
Affected area 15
Affected area 16
Affected area 17
Affected area 18
Affected area 19
Affected area 20
Affected area 21
Affected area 22
Affected area 23
Company & contacts
Company & contacts
Brand: Ortho-clinical Diagnostics. Contact: Ortho-Clinical Diagnostics.
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.