Open official notice

StealthStation Cranial Software, Model: 9735585, used with StealthStation S7 i7 systems

Official notice

14 Jun 2023

RecallScope risk estimate:CriticalFDA

StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

556

Recall number

Z-1666-2023

Official notice

2023-06-14

The hazard

In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

92269

Recall Numbers

Z-1666-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems

Units affected

556

Hazard

In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

BG · Bulgaria

Affected area 8

BR · Brazil

Affected area 9

CA · Canada

Affected area 10

CH · Switzerland

Affected area 11

CL · Chile

Affected area 12

CO · Colombia

Affected area 13

CZ · Czech Republic

Affected area 14

DE · Germany

Affected area 15

DK · Denmark

Affected area 16

DO · Dominican Republic

Affected area 17

ES · Spain

Affected area 18

FI · Finland

Affected area 19

FR · France

Affected area 20

GR · Greece

Affected area 21

HR · Croatia

Affected area 22

HU · Hungary

Affected area 23

IE · Ireland

Affected area 24

IL · Israel

Affected area 25

IN · India

Affected area 26

IT · Italy

Affected area 27

JP · Japan

Affected area 28

KR · South Korea

Affected area 29

LV · Latvia

Affected area 30

MX · Mexico

Affected area 31

NL · Netherlands

Affected area 32

PK · Pakistan

Affected area 33

PL · Poland

Affected area 34

PT · Portugal

Affected area 35

RO · Romania

Affected area 36

SA · Saudi Arabia

Affected area 37

SE · Sweden

Affected area 38

SG · Singapore

Affected area 39

SI · Slovenia

Affected area 40

TH · Thailand

Affected area 41

TR · Turkey

Affected area 42

ZA · South Africa

Company & contacts

Manufacturer

Medtronic Navigation, Inc.

Phone

Not provided

Brand: Medtronic Navigation. Contact: Medtronic Navigation, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-1666-2023

Event ID

92269

Units affected

556

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Jun 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button