Image not available in RecallScope.
Persona Cemented Tibial Broach Inserter Extractor. The Persona Cemented Tibial Broach Inserter Extractor Handle is a Use sterilizable
Official notice
31 Dec 2014
An official FDA notice concerns Persona Cemented Tibial Broach Inserter Extractor. The Persona Cemented Tibial Broach Inserter Extractor Handle is a Use sterilizable. The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi Published 2014-12-31.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
83
Recall number
Z-0853-2015
Official notice
2014-12-31
The hazard
The hazard
The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Lots
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Location 2
Location 3
Location 4
Location 5
Location 6
Location 7
Location 8
Location 9
Location 10
Location 11
Location 12
Affected area 1
Affected area 2
Affected area 3
Affected area 4
Affected area 5
Affected area 6
Company & contacts
Company & contacts
Brand: Zimmer. Contact: Zimmer, Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.