Nextgen Complete Knee Solution Femoral and Provisional Impactor Extractor, REF 00-5901-026-00, Non-Sterile Zimmer, Warsaw Indiana The

Official notice

8 Aug 2012

RecallScope risk estimate:HighFDA

An official FDA notice concerns Nextgen Complete Knee Solution Femoral and Provisional Impactor Extractor, REF 00-5901-026-00, Non-Sterile Zimmer, Warsaw Indiana The. The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall Published 2012-08-08.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

1,662

Recall number

Z-2091-2012

Official notice

2012-08-08

The hazard

The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

11002148

Event Ids

61764

Recall Numbers

Z-2091-2012

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Nextgen Complete Knee Solution Femoral and Provisional Impactor/ Extractor, REF 00-5901-026-00, Non-Sterile Zimmer, Warsaw Indiana The instrument is attached to the femoral implant or provisional to aid in positioning and inserting the components.

Units affected

1,662

Hazard

The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Affected area 4

CN · China

Affected area 5

DE · Germany

Affected area 6

FI · Finland

Affected area 7

HK · Hong Kong

Affected area 8

IN · India

Affected area 9

JP · Japan

Affected area 10

SG · Singapore

Company & contacts

Brand

Manufacturer

Zimmer Inc.

Phone

Not provided

Brand: Zimmer. Contact: Zimmer Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2091-2012

Event ID

61764

Units affected

1,662

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 8 Aug 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button