Integra CRW Precision Arc (CRWPRECISE), CRW Arc System (CRWASL), HRAIM Head Ring Assembly with Intubation Mounts The CRW stereotactic

Official notice

26 Sept 2012

RecallScope risk estimate:MediumFDA

Integra CRW Precision Arc (CRWPRECISE), CRW Arc System (CRWASL), HRAIM Head Ring Assembly with Intubation Mounts The CRW stereotactic system is used in conjunction with the HRAIM head ring is a target-centered system that can be configured to be CT-only or CTIMR compatible. The CRW stereotactic system is a multi-purpose system used for localizing intercranial targets for precisely directing instruments such as: Biopsy forceps RF lesioning electrodes Deep brain electrodes Recording and stimulating electrodes

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

175

Recall number

Z-2428-2012

Official notice

2012-09-26

The hazard

T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened, the T-handle could stop in a vertical position that prevents the CRWPRECISE and certain models of the CRW-ASL from seating properly on the head ring although it appears to be fully seated

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

62978

Recall Numbers

Z-2428-2012

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Integra CRW Precision Arc (CRWPRECISE), CRW Arc System (CRWASL), HRAIM Head Ring Assembly with Intubation Mounts The CRW stereotactic system is used in conjunction with the HRAIM head ring is a target-centered system that can be configured to be CT-only or CTIMR compatible. The CRW stereotactic system is a multi-purpose system used for localizing intercranial targets for precisely directing instruments such as: Biopsy forceps RF lesioning electrodes Deep brain electrodes Recording and stimulating electrodes

Units affected

175

Hazard

T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened, the T-handle could stop in a vertical position that prevents the CRWPRECISE and certain models of the CRW-ASL from seating properly on the head ring although it appears to be fully seated

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AU · Australia

Affected area 4

BE · Belgium

Affected area 5

CA · Canada

Affected area 6

CN · China

Affected area 7

DK · Denmark

Affected area 8

ES · Spain

Affected area 9

FR · France

Affected area 10

GB · United Kingdom

Affected area 11

HK · Hong Kong

Affected area 12

HU · Hungary

Affected area 13

IT · Italy

Affected area 14

JP · Japan

Affected area 15

MY · Malaysia

Affected area 16

SG · Singapore

Affected area 17

ZA · South Africa

Company & contacts

Manufacturer

Integra LifeSciences Corporation

Phone

Not provided

Brand: Integra Lifesciences. Contact: Integra LifeSciences Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2428-2012

Event ID

62978

Units affected

175

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 26 Sept 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button