Open official notice

Alere INRatio PT INR System Professional Model Numbers: 0100004 Product Usage: Alere INRatio ¿ 2 PT INR Monitoring System (Professional

Official notice

31 Aug 2016

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Alere INRatio PT INR System Professional Model Numbers: 0100004 Product Usage: Alere INRatio ¿ 2 PT INR Monitoring System (Professional. Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory. Published 2016-08-31.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Recall number

Z-2354-2016

Official notice

2016-08-31

The hazard

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

74665

Recall Numbers

Z-2354-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Alere INRatio PT/INR System Professional Model Numbers: 0100004 Product Usage: Alere INRatio ¿ 2 PT/INR Monitoring System (Professional Use): The Alere INRatio ¿ 2 PT/INR Monitoring System (Professional Use), consisting of the INRatio ¿ 2 Monitor and INRatio¿ 2 PT/INR test strip, is used for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin on clotting time by health care professionals. The Alere INRatio¿ 2 PT/INR Monitoring System (Professional Use) is intended for use outside of the body (in vitro diagnostic use). The Alere INRatio¿ 2 PT/INR Monitoring System (Professional Use) is not intended to be used for screening purposes.

Hazard

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AU · Australia

Affected area 4

BE · Belgium

Affected area 5

BR · Brazil

Affected area 6

CA · Canada

Affected area 7

CL · Chile

Affected area 8

CN · China

Affected area 9

CO · Colombia

Affected area 10

EG · Egypt

Affected area 11

ES · Spain

Affected area 12

FR · France

Affected area 13

HK · Hong Kong

Affected area 14

IL · Israel

Affected area 15

IN · India

Affected area 16

IT · Italy

Affected area 17

JP · Japan

Affected area 18

KR · South Korea

Affected area 19

MY · Malaysia

Affected area 20

NL · Netherlands

Affected area 21

RU · Russia

Affected area 22

SG · Singapore

Affected area 23

TH · Thailand

Affected area 24

TW · Taiwan

Affected area 25

UA · Ukraine

Affected area 26

UY · Uruguay

Affected area 27

VN · Vietnam

Affected area 28

ZA · South Africa

Company & contacts

Manufacturer

Alere San Diego, Inc.

Phone

Not provided

Brand: Alere San Diego. Contact: Alere San Diego, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-2354-2016

Event ID

74665

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 31 Aug 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button