A610 Clinician Programmer Application for Deep Brain Stimulation Clinician Programming Guide Percept PC Model B35200 Neurostimulator

Official notice

14 Apr 2021

RecallScope risk estimate:MediumFDA

An official FDA notice concerns A610 Clinician Programmer Application for Deep Brain Stimulation Clinician Programming Guide Percept PC Model B35200 Neurostimulator. There are two issues within this correction action: 1. Cycling Issue: When copying settings from a previously used Activa device to a new Percept PC Implantable Neurostimulator (INS) using the "replacement" function of the A610 clinician programmer application, the cycling feature is not copied properly. 2. Log Fill Issue: When the Percept PC INS System Event log is full, the clinician programmer A610 Software application version 2.0.4584, 2.0.4594, 2.0.4605, or 3.0.1057 crashes when it attempts a reading of the last records in the log. Published 2021-04-14.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

5,332

Recall number

Z-1368-2021

Official notice

2021-04-14

The hazard

There are two issues within this correction action: 1. Cycling Issue: When copying settings from a previously used Activa device to a new Percept PC Implantable Neurostimulator (INS) using the "replacement" function of the A610 clinician programmer application, the cycling feature is not copied properly. 2. Log Fill Issue: When the Percept PC INS System Event log is full, the clinician programmer A610 Software application version 2.0.4584, 2.0.4594, 2.0.4605, or 3.0.1057 crashes when it attempts a reading of the last records in the log.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

87569

Recall Numbers

Z-1368-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

A610 Clinician Programmer Application for Deep Brain Stimulation Clinician Programming Guide Percept PC Model B35200 Neurostimulator Application, version 2.0.4584, 2.0.4594, 2.0.4605, and 3.0.1057* * Version 3.0.1057 is approved for distribution in Europe only.

Units affected

5,332

Hazard

There are two issues within this correction action: 1. Cycling Issue: When copying settings from a previously used Activa device to a new Percept PC Implantable Neurostimulator (INS) using the "replacement" function of the A610 clinician programmer application, the cycling feature is not copied properly. 2. Log Fill Issue: When the Percept PC INS System Event log is full, the clinician programmer A610 Software application version 2.0.4584, 2.0.4594, 2.0.4605, or 3.0.1057 crashes when it attempts a reading of the last records in the log.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AT · Austria

Affected area 3

AU · Australia

Affected area 4

BE · Belgium

Affected area 5

BG · Bulgaria

Affected area 6

BR · Brazil

Affected area 7

CA · Canada

Affected area 8

CH · Switzerland

Affected area 9

CZ · Czech Republic

Affected area 10

DE · Germany

Affected area 11

DK · Denmark

Affected area 12

ES · Spain

Affected area 13

FI · Finland

Affected area 14

FR · France

Affected area 15

GB · United Kingdom

Affected area 16

GR · Greece

Affected area 17

HR · Croatia

Affected area 18

HU · Hungary

Affected area 19

IE · Ireland

Affected area 20

IL · Israel

Affected area 21

IS · Iceland

Affected area 22

IT · Italy

Affected area 23

JP · Japan

Affected area 24

LT · Lithuania

Affected area 25

LU · Luxembourg

Affected area 26

MT · Malta

Affected area 27

MY · Malaysia

Affected area 28

NL · Netherlands

Affected area 29

NO · Norway

Affected area 30

PL · Poland

Affected area 31

PT · Portugal

Affected area 32

SA · Saudi Arabia

Affected area 33

SE · Sweden

Affected area 34

SI · Slovenia

Affected area 35

SK · Slovakia

Affected area 36

TH · Thailand

Affected area 37

TR · Turkey

Company & contacts

Manufacturer

Medtronic Neuromodulation

Phone

Not provided

Brand: Medtronic Neuromodulation. Contact: Medtronic Neuromodulation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1368-2021

Event ID

87569

Units affected

5,332

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Apr 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button