No image published for this recall

No image published for this recall

Yukon lock shaft inserter

Official notice

13 Jun 2022

Last checked

13 Jun 2022

Medium riskHealth Canada

An official Health Canada notice concerns Yukon lock shaft inserter. Stryker received reports of Yukon Lock Shaft Inserters, lot MXGD, having difficulty threading into screws and/or thread deformation observed on the inserter. Further investigation identified the thread deformation to be the result of a manufacturing non-conformance. No adverse events have been reported for this issue. Recall start date: June 3, 2022 Published 2022-06-13.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64259

Official notice

2022-06-13

The hazard

Stryker received reports of Yukon Lock Shaft Inserters, lot MXGD, having difficulty threading into screws and/or thread deformation observed on the inserter. Further investigation identified the thread deformation to be the result of a manufacturing non-conformance. No adverse events have been reported for this issue. Recall start date: June 3, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Lots

MXGD, having

Recall Numbers

64259

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Yukon Lock Shaft Inserter, Lot or serial number - MXGD-444564, Model or catalogue number - 7601-90186

Risk / hazard

Stryker received reports of Yukon Lock Shaft Inserters, lot MXGD, having difficulty threading into screws and/or thread deformation observed on the inserter. Further investigation identified the thread deformation to be the result of a manufacturing non-conformance. No adverse events have been reported for this issue. Recall start date: June 3, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: K2m, Yukon Lock Shaft Inserter.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

64259

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Straight rod - ti

Medium riskHealth Canada20 Jul 2022
CA
No image published for this recall

Intramedullary bone saw blade assembly

Medium riskHealth Canada3 Jul 2026
CA
No image published for this recall

Glidesheath slender® introducer sheath

Medium riskHealth Canada2 Jul 2026
CA
Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)Image: Health Canada recall notice

Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Medium riskHealth Canada22 Jun 2026
CA
No image published for this recall

Penumbra system red 68 reperfusion catheter

Critical riskHealth Canada22 Jun 2026
CA
No image published for this recall

Faradrive™ steerable sheath

Medium riskHealth Canada19 Jun 2026
CA
No image published for this recall

Harmony™ delivery catheter system

Medium riskHealth Canada4 Jun 2026
CA
No image published for this recall

Monnal™ t60 advanced

Medium riskHealth Canada30 Apr 2026
CA
No image published for this recall

Medtronic dlp retrograde cardioplegia cannulae

Medium riskHealth Canada30 Apr 2026
CA
No image published for this recall

Cook medical beacon® tip centimeter sizing catheter

Medium riskHealth Canada28 Apr 2026
CA
No image published for this recall

Alecensaro (alectinib) and the risk of severe hypertriglyceridemia

Critical riskHealth Canada22 Apr 2026
CA
No image published for this recall

Synchromed clinician programmer application

Critical riskHealth Canada15 Apr 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 13 Jun 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.