Video telescope - endoeye hd ii
Official notice
27 Nov 2023
Last checked
27 Nov 2023
An official Health Canada notice concerns Video telescope - endoeye hd ii. Olympus has identified a total of 1003 complaints, including 372 adverse events (since September 2020 to August 2023) related to pink or green coloration of the image, including cases with reported delays of treatments and/or prolonged surgery. Olympus investigated the complaints and identified that the CCD chip that produces the image became damaged. The CCD chip is sensitive to heat and mechanical shocks. Olympus's risk management is taking this into account, and the instructions for use mention necessary steps to prepare the video telescope prior to usage. This medical device recall has been launched to remind customers to follow the steps in the instructions for use, especially to inspect the image prior to a clinical procedure, and always have a spare laparoscope available. Recall start date: November 8, 2023 Published 2023-11-27.
What to do now
Contact the manufacturer if you require additional information.
Recall number
74656
Official notice
2023-11-27
The hazard
The hazard
Olympus has identified a total of 1003 complaints, including 372 adverse events (since September 2020 to August 2023) related to pink or green coloration of the image, including cases with reported delays of treatments and/or prolonged surgery. Olympus investigated the complaints and identified that the CCD chip that produces the image became damaged. The CCD chip is sensitive to heat and mechanical shocks. Olympus's risk management is taking this into account, and the instructions for use mention necessary steps to prepare the video telescope prior to usage. This medical device recall has been launched to remind customers to follow the steps in the instructions for use, especially to inspect the image prior to a clinical procedure, and always have a spare laparoscope available. Recall start date: November 8, 2023
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Olympus Winter Ibe, Video Telescope - Endoeye HD Ii.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Olympus hf cable

Celon probreath and celon prosleep applicators

Oes pro urology resection sheaths

Olympus resection and inner sheaths

Electrosurgical generator, esg-410

Oes pro urology resection sheaths

Electrosurgical generator, esg-410

Rapidlab® 348ex system

Airseal access port with blunt obturator and airseal access port with optical obturator

Wasplab

Nephromax high pressure nephrostomy balloon catheter

Uhi-4 high flow insufflation system - insufflator
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 27 Nov 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.