Sonicaid freedom wireless fetal monitoring system

Official notice

26 Oct 2022

Last checked

26 Oct 2022

Medium riskHealth Canada

An official Health Canada notice concerns Sonicaid freedom wireless fetal monitoring system. Huntleigh has received a report from a user regarding the potential for inappropriate clinical diagnosis and action due to the incorrect setup of the RF channels. In the particular case in question, both units had been inadvertently set to the same RF channel and were placed in close proximity to each other, leading to the second unit displaying the FHR from the first unit. Recall start date: October 19, 2022 Published 2022-10-26.

What to do now

Contact the manufacturer if you require additional information.

Recall number

72082

Official notice

2022-10-26

The hazard

Huntleigh has received a report from a user regarding the potential for inappropriate clinical diagnosis and action due to the incorrect setup of the RF channels. In the particular case in question, both units had been inadvertently set to the same RF channel and were placed in close proximity to each other, leading to the second unit displaying the FHR from the first unit. Recall start date: October 19, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

72082

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Sonicaid Freedom Wireless Fetal Monitoring System, Lot or serial number - 778EW10002421 778EW10002621 778EW10002521, Model or catalog number - SF1-SL

Risk / hazard

Huntleigh has received a report from a user regarding the potential for inappropriate clinical diagnosis and action due to the incorrect setup of the RF channels. In the particular case in question, both units had been inadvertently set to the same RF channel and were placed in close proximity to each other, leading to the second unit displaying the FHR from the first unit. Recall start date: October 19, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Huntleigh Healthcare, Sonicaid Freedom Wireless Fetal Monitoring System.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

72082

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Team3 intrapartum and team3 antepartum

Medium riskHealth Canada18 May 2023
CA
No image published for this recall

Team3 intrapartum and team3 antepartum

Medium riskHealth Canada13 Apr 2023
CA
Vevor 4 in 1 baby walker recalledImage: Health Canada recall notice

Vevor 4 in 1 baby walker recalled

Critical riskHealth Canada6 Jul 2026
CA
Amana through-the-wall air conditioners or heat pumps recalledImage: Health Canada recall notice

Amana through-the-wall air conditioners or heat pumps recalled

High riskHealth Canada25 Jun 2026
CA
No image published for this recall

A.t.s® 4000 ts tourniquet systems single and dual hose with cpc connectors

Medium riskHealth Canada23 Jun 2026
CA
No image published for this recall

Iceman classic3™

Medium riskHealth Canada3 Jun 2026
CA
Generac power electric start pressure washers recalledImage: Health Canada recall notice

Generac power electric start pressure washers recalled

Critical riskHealth Canada14 May 2026
CA
Zwilling enfinigy electric kettle and zwilling enfinigy electric kettle pro recalledImage: Health Canada recall notice

Zwilling enfinigy electric kettle and zwilling enfinigy electric kettle pro recalled

Medium riskHealth Canada14 May 2026
CA
Fridababy canadian model (water-filled) not-too-cold-to-hold teethers recalledImage: Health Canada recall notice

Fridababy canadian model (water-filled) not-too-cold-to-hold teethers recalled

Medium riskHealth Canada24 Apr 2026
CA
Ecosmart fire stix portable fire pit and mini t designer fireplace recalledImage: Health Canada recall notice

Ecosmart fire stix portable fire pit and mini t designer fireplace recalled

Critical riskHealth Canada21 Apr 2026
CA
Generac portable generators recalledImage: Health Canada recall notice

Generac portable generators recalled

High riskHealth Canada16 Apr 2026
CA
No image published for this recall

I‑stat eg7+ cartridge

Critical riskHealth Canada15 Apr 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 26 Oct 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.