Siemens - Mammomat inspiration
Official notice
24 Oct 2025
Last checked
24 Oct 2025
An official Health Canada notice concerns Siemens - Mammomat inspiration. The operator table (part number 14467741) is not intended and not released for bus installations as it was designed for stationary operation only. Due to the impacts and/or vibrations during transportation, there is a risk that parts of the operator table (e.g. screws, monitor, monitor holder) could come loose and potentially fall off resulting in a risk of serious injury to the operating staff from falling parts. In addition, damage to the lead shield may also occur. Published 2025-10-24.
What to do now
Contact the manufacturer if you require additional information.
Official notice
2025-10-24
The hazard
The hazard
The operator table (part number 14467741) is not intended and not released for bus installations as it was designed for stationary operation only. Due to the impacts and/or vibrations during transportation, there is a risk that parts of the operator table (e.g. screws, monitor, monitor holder) could come loose and potentially fall off resulting in a risk of serious injury to the operating staff from falling parts. In addition, damage to the lead shield may also occur.
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Siemens.
Source & status
Source & status
Source
Tier
Category
App estimate
Recall details
Recall details
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Rapidlab® 348ex system

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Artis icono ceiling

Artis icono ceiling, floor and biplane systems

Jeep - Fiat chrysler automobiles

Lincoln - Ford

Dimension® creatinine (cre2)
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 24 Oct 2025. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.