Novum iq lvp

Official notice

4 Jan 2022

Last checked

4 Jan 2022

Medium riskHealth Canada

An official Health Canada notice concerns Novum iq lvp. Baxter Corporation has internally identified a scenario on the Novum IQ Large Volume Infusion Pump (lVP) where 1) an unintended second bolus or 2) an unintended second auto-programming order update may be delivered to the patient. Baxter Corporation has identified an additional scenario where a rapid double key press will result in unintended behavior on the Novum IQ LVP. This issue will occur if the user rapidly double presses the 'OK' key (<200 milliseconds apart) when accepting a value exceeding the drug library soft limit. Recall start date: December 29, 2021. Published 2022-01-04.

What to do now

Contact the manufacturer if you require additional information.

Recall number

63792

Official notice

2022-01-04

The hazard

Baxter Corporation has internally identified a scenario on the Novum IQ Large Volume Infusion Pump (lVP) where 1) an unintended second bolus or 2) an unintended second auto-programming order update may be delivered to the patient. Baxter Corporation has identified an additional scenario where a rapid double key press will result in unintended behavior on the Novum IQ LVP. This issue will occur if the user rapidly double presses the 'OK' key (<200 milliseconds apart) when accepting a value exceeding the drug library soft limit. Recall start date: December 29, 2021.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

63792

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected Products - Novum Iq Lvp, Lot or serial number - All lots., Model or catalog number - 40700BAX

Risk / hazard

Baxter Corporation has internally identified a scenario on the Novum IQ Large Volume Infusion Pump (lVP) where 1) an unintended second bolus or 2) an unintended second auto-programming order update may be delivered to the patient. Baxter Corporation has identified an additional scenario where a rapid double key press will result in unintended behavior on the Novum IQ LVP. This issue will occur if the user rapidly double presses the 'OK' key (<200 milliseconds apart) when accepting a value exceeding the drug library soft limit. Recall start date: December 29, 2021.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Baxter Healthcare, Novum Iq Lvp.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

63792

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Novum iq lvp

Medium riskHealth Canada14 Aug 2025
CA
No image published for this recall

Novum iq lvp

Medium riskHealth Canada22 Jul 2025
CA
No image published for this recall

Novum iq lvp

Medium riskHealth Canada14 May 2025
CA
No image published for this recall

Clearlink iv sets

Medium riskHealth Canada16 Sept 2025
CA
No image published for this recall

Wireless battery module

Medium riskHealth Canada1 Aug 2024
CA
No image published for this recall

Novum iq lvp

Medium riskHealth Canada20 Jun 2024
CA
No image published for this recall

Baxter healthcare hemostatic products

Medium riskHealth Canada1 Dec 2023
CA
No image published for this recall

Novum iq large volume pump operators manual

Medium riskHealth Canada8 Nov 2023
CA
No image published for this recall

Novum iq large volume pump unapproved adhesive

Medium riskHealth Canada8 Nov 2023
CA
No image published for this recall

Novum iq large volume pump keypads

Medium riskHealth Canada7 Nov 2023
CA
No image published for this recall

Novum iq lvp

High riskHealth Canada14 Aug 2023
CA
No image published for this recall

Novum iq large volume pump

High riskHealth Canada2 Aug 2023

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 4 Jan 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.