Mcgrath™ mac video laryngoscope

Official notice

23 Jul 2024

CriticalHealth Canada

An official Health Canada notice concerns Mcgrath™ mac video laryngoscope. Medtronic is issuing a voluntary recall for item codes 300-000-000 and 300-200-000 (section 1). This notice also includes information related to an addendum to the instructions for use (IFU) for item code 301-000-000 (section 2). Section 1 (for item codes 300-000-000 and 300-200-000) : Customers should immediately discontinue use of these devices that are being recalled. Customers should uninstall the battery assembly from these devices, dispose as per local procedure and follow manufacturer's instructions for returning the devices that are being recalled. To date, manufacturer has determined that the battery management system within the devices being recalled is not sufficient to mitigate against depletion of the battery voltage below the design threshold. Batteries that are depleted below the design threshold may increase the risk of the battery becoming unstable, which may lead to a thermal event followed by a risk of explosion. A thermal event followed by a risk of explosion of the battery assembly can potentially cause caregivers and/or patients to experience burns, lacerations, tinnitus, hearing impairment and acoustic shock. The potential for a delay to treatment, respiratory failure, hypoxia, unspecified tissue injury, scar tissue, foreign body in patient or caregiver, tooth loss, and eye injury also exist. There also exists the potential for damage to surrounding equipment or surfaces. This notice follows the receipt of two customer reports stating that the installed battery experienced a thermal event, which resulted in an explosion of the battery assembly before patient use. In one of these two instances, there were caregiver injuries. Section 2 (for item code 301-000-000) : Item code 301-000-000 are not being removed and are safe and effective for patient use. Next-generation laryngoscopes include an advanced battery management system that maintains the battery within intended voltage usage levels and renders them inoperable when the battery voltage falls below the device design threshold. Manufacturer is issuing an addendum to the instructions for use (IFU). This addendum will include the following warnings regarding proper battery handling. Improper battery handling can increase the risk of a thermal event. Battery assembly (340-000-000) is labelled with a "use by date". Do not use a battery past the "use by date". Refer to the product instructions for use (IFU) for proper battery storage conditions. Any battery assemblies that have not been stored in these conditions should be disposed of per local guidelines. Do not use a battery assembly, installed or uninstalled, that has been dropped as it may have sustained internal damage that is not visible. Dispose of any dropped battery assembly in accordance with facility's procedure and replace with a new labeled battery assembly. Recall start date: July 18, 2024 Published 2024-07-23.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75854

Official notice

2024-07-23

The hazard

Medtronic is issuing a voluntary recall for item codes 300-000-000 and 300-200-000 (section 1). This notice also includes information related to an addendum to the instructions for use (IFU) for item code 301-000-000 (section 2). Section 1 (for item codes 300-000-000 and 300-200-000) : Customers should immediately discontinue use of these devices that are being recalled. Customers should uninstall the battery assembly from these devices, dispose as per local procedure and follow manufacturer's instructions for returning the devices that are being recalled. To date, manufacturer has determined that the battery management system within the devices being recalled is not sufficient to mitigate against depletion of the battery voltage below the design threshold. Batteries that are depleted below the design threshold may increase the risk of the battery becoming unstable, which may lead to a thermal event followed by a risk of explosion. A thermal event followed by a risk of explosion of the battery assembly can potentially cause caregivers and/or patients to experience burns, lacerations, tinnitus, hearing impairment and acoustic shock. The potential for a delay to treatment, respiratory failure, hypoxia, unspecified tissue injury, scar tissue, foreign body in patient or caregiver, tooth loss, and eye injury also exist. There also exists the potential for damage to surrounding equipment or surfaces. This notice follows the receipt of two customer reports stating that the installed battery experienced a thermal event, which resulted in an explosion of the battery assembly before patient use. In one of these two instances, there were caregiver injuries. Section 2 (for item code 301-000-000) : Item code 301-000-000 are not being removed and are safe and effective for patient use. Next-generation laryngoscopes include an advanced battery management system that maintains the battery within intended voltage usage levels and renders them inoperable when the battery voltage falls below the device design threshold. Manufacturer is issuing an addendum to the instructions for use (IFU). This addendum will include the following warnings regarding proper battery handling. Improper battery handling can increase the risk of a thermal event. Battery assembly (340-000-000) is labelled with a "use by date". Do not use a battery past the "use by date". Refer to the product instructions for use (IFU) for proper battery storage conditions. Any battery assemblies that have not been stored in these conditions should be disposed of per local guidelines. Do not use a battery assembly, installed or uninstalled, that has been dropped as it may have sustained internal damage that is not visible. Dispose of any dropped battery assembly in accordance with facility's procedure and replace with a new labeled battery assembly. Recall start date: July 18, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75854

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - McGRATH™ MAC Video Laryngoscope, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - 300-000-000 301-000-000

Product 2

Affected products - McGRATH™ MAC EMS Video Laryngoscope, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - 300-200-000

Risk / hazard

Medtronic is issuing a voluntary recall for item codes 300-000-000 and 300-200-000 (section 1). This notice also includes information related to an addendum to the instructions for use (IFU) for item code 301-000-000 (section 2). Section 1 (for item codes 300-000-000 and 300-200-000) : Customers should immediately discontinue use of these devices that are being recalled. Customers should uninstall the battery assembly from these devices, dispose as per local procedure and follow manufacturer's instructions for returning the devices that are being recalled. To date, manufacturer has determined that the battery management system within the devices being recalled is not sufficient to mitigate against depletion of the battery voltage below the design threshold. Batteries that are depleted below the design threshold may increase the risk of the battery becoming unstable, which may lead to a thermal event followed by a risk of explosion. A thermal event followed by a risk of explosion of the battery assembly can potentially cause caregivers and/or patients to experience burns, lacerations, tinnitus, hearing impairment and acoustic shock. The potential for a delay to treatment, respiratory failure, hypoxia, unspecified tissue injury, scar tissue, foreign body in patient or caregiver, tooth loss, and eye injury also exist. There also exists the potential for damage to surrounding equipment or surfaces. This notice follows the receipt of two customer reports stating that the installed battery experienced a thermal event, which resulted in an explosion of the battery assembly before patient use. In one of these two instances, there were caregiver injuries. Section 2 (for item code 301-000-000) : Item code 301-000-000 are not being removed and are safe and effective for patient use. Next-generation laryngoscopes include an advanced battery management system that maintains the battery within intended voltage usage levels and renders them inoperable when the battery voltage falls below the device design threshold. Manufacturer is issuing an addendum to the instructions for use (IFU). This addendum will include the following warnings regarding proper battery handling. Improper battery handling can increase the risk of a thermal event. Battery assembly (340-000-000) is labelled with a "use by date". Do not use a battery past the "use by date". Refer to the product instructions for use (IFU) for proper battery storage conditions. Any battery assemblies that have not been stored in these conditions should be disposed of per local guidelines. Do not use a battery assembly, installed or uninstalled, that has been dropped as it may have sustained internal damage that is not visible. Dispose of any dropped battery assembly in accordance with facility's procedure and replace with a new labeled battery assembly. Recall start date: July 18, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Covidien, Mcgrath™ Mac Video Laryngoscope.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type I

App estimate

critical

Recall details

Recall number

75854

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

HOME
High

Endoflip system

Health Canada14 May 2026

Brand match

High risk

Markets

CA
HOME
High

Polysorb coated braided absorbable single stitch reload for use with the endo stitch and sils stitch suturing devices

Health Canada24 Apr 2026

Brand match

High risk

Markets

CA
HOME
Medium

Bipolar hugo ras systems

Health Canada19 Feb 2026

Brand match

Medium risk

Markets

CA
HOME
Medium

Hugo ras tower, 120v

Health Canada9 May 2025

Brand match

Medium risk

Markets

CA
HOME
High

Esoflip dilation catheter 30mm

Health Canada3 May 2024

Brand match

High risk

Markets

CA
HOME
Medium

Hugo ras tower, 120v

Health Canada3 Apr 2024

Brand match

Medium risk

Markets

CA
HOME
Critical

Hugo ras surgeon console

Health Canada2 Apr 2024

Brand match

Critical risk

Markets

CA
HOME
Medium

Auto suture blunt tip and structural balloon trocar

Health Canada13 Mar 2024

Brand match

Medium risk

Markets

CA
HOME
High

Sofsilk™ braided silk sutures and ti-cron™ coated braided polyester sutures

Health Canada31 Jan 2024

Brand match

High risk

Markets

CA
HOME
Medium

Valleylab ft10 ft series energy platform

Health Canada5 Dec 2023

Brand match

Medium risk

Markets

CA
HOME
Medium

Hugo™ robotic-assisted surgery (ras) tower, 120v

Health Canada15 Sept 2023

Brand match

Medium risk

Markets

CA
HOME
High

Mahurkar acute high pressure triple lumen catheter kit curved extensions

Health Canada7 Jul 2023

Brand match

High risk

Markets

CA

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 23 Jul 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.