Epoc® nxs host – 5553ca

Official notice

17 Nov 2023

Last checked

17 Nov 2023

Medium riskHealth Canada

An official Health Canada notice concerns Epoc® nxs host – 5553ca. Siemens Healthcare Diagnostics Inc. Has confirmed a potential issue that could occur under specific conditions when the (automatic) delete blood tests feature is enabled. When the results are run on the epoc® NXS host and delete blood tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. Recall start date: October 24, 2023 Published 2023-11-17.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74607

Official notice

2023-11-17

The hazard

Siemens Healthcare Diagnostics Inc. Has confirmed a potential issue that could occur under specific conditions when the (automatic) delete blood tests feature is enabled. When the results are run on the epoc® NXS host and delete blood tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. Recall start date: October 24, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

74607

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - epoc® NXS Host – 5553CA, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - 11413506

Risk / hazard

Siemens Healthcare Diagnostics Inc. Has confirmed a potential issue that could occur under specific conditions when the (automatic) delete blood tests feature is enabled. When the results are run on the epoc® NXS host and delete blood tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. Recall start date: October 24, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Epocal, Epoc® Nxs Host – 5553ca.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

74607

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Epoc nxs host - 5553ca

Medium riskHealth Canada12 Dec 2024
CA
No image published for this recall

Heartsine® samaritan® pad sam 350p sam 360p sam 500p - membrane tail

Medium riskHealth Canada23 Jul 2026
CA
No image published for this recall

Aisys, avance, and aespire anesthesia machines

Medium riskHealth Canada2 Jul 2026
CA
No image published for this recall

Abiomed introducer kits with impella heart pumps

High riskHealth Canada29 Jun 2026
CA
Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)Image: Health Canada recall notice

Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Medium riskHealth Canada22 Jun 2026
CA
No image published for this recall

Penumbra system red 68 reperfusion catheter

Critical riskHealth Canada22 Jun 2026
CA
No image published for this recall

Ge healthcare integrated and stand-alone resuscitation systems

Medium riskHealth Canada16 Jun 2026
CA
No image published for this recall

Unauthorized soft-shelled hyperbaric chambers may pose serious health risks 2026-06-16

Critical riskHealth Canada16 Jun 2026
CA
No image published for this recall

Etms 70 mm cooled coil

Medium riskHealth Canada11 Jun 2026
CA
No image published for this recall

Giraffe™ omnibed™ carestation™

Medium riskHealth Canada4 Jun 2026
CA
No image published for this recall

Harmony™ delivery catheter system

Medium riskHealth Canada4 Jun 2026
CA
No image published for this recall

Omnipod® insulin management system - omnipod®

Critical riskHealth Canada1 Jun 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 17 Nov 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.