No image published for this recall
Durasul modular head
Official notice
15 Mar 2023
Last checked
15 Mar 2023
An official Health Canada notice concerns Durasul modular head. Zimmer Biomet is conducting a medical device correction to update the compatibility matrix as referred to in the instructions for use (IFU) for the CoCr Femoral Head XS. The update is to remove the compatibility with the Epsilon Durasul Constrained Acetabular liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009). When used in combination with the Epsilon Durasul Constrained Acetabular liner, the CoCr Femoral Head XS has a range of motion of 76 degrees. The reduced range of motion could potentially result in the health risks identified. There have been no complaints for this issue. Recall start date: March 3, 2023 Published 2023-03-15.
What to do now
Contact the manufacturer if you require additional information.
Recall number
73329
Official notice
2023-03-15
The hazard
The hazard
Zimmer Biomet is conducting a medical device correction to update the compatibility matrix as referred to in the instructions for use (IFU) for the CoCr Femoral Head XS. The update is to remove the compatibility with the Epsilon Durasul Constrained Acetabular liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009). When used in combination with the Epsilon Durasul Constrained Acetabular liner, the CoCr Femoral Head XS has a range of motion of 76 degrees. The reduced range of motion could potentially result in the health risks identified. There have been no complaints for this issue. Recall start date: March 3, 2023
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Zimmer, Durasul Modular Head.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Mixing bowl and spatula, 20- pack

Dual mobility vivacit-e® vitamin e bearing

Cpt hip system femoral stem 12 14 neck taper

Nexgen cr-flex femoral component, size e and size e minus (e-)

Zplp distal lateral fibular plate

Zimmer m l taper taperfemoral stem

Nexgen complete knee solution stemmed non-augment tibial comp (cr ps) option

Ncb screw 5.0 l=38

Heartsine® samaritan® pad sam 350p sam 360p sam 500p - membrane tail

Intramedullary bone saw blade assembly

Glidesheath slender® introducer sheath

Abiomed introducer kits with impella heart pumps
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 15 Mar 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.