Chromophare surgical lighting system

Official notice

5 Mar 2026

Last checked

5 Mar 2026

Medium riskHealth Canada
Watch Stryker Communications

An official Health Canada notice concerns Chromophare surgical lighting system. Stryker has discovered that certain lots of the soffit ring may not adequately support the weight of the ceiling cover. As a result, the cover may shift or slide down the ceiling tube. Movement of the ceiling cover can expose the area above the ceiling, which may allow particles, such as dust, debris, residue, or other contaminants, to enter the sterile field. Published 2026-03-05.

What to do now

Contact the manufacturer if you require additional information.

Recall number

81701

Official notice

2026-03-05

The hazard

Stryker has discovered that certain lots of the soffit ring may not adequately support the weight of the ceiling cover. As a result, the cover may shift or slide down the ceiling tube. Movement of the ceiling cover can expose the area above the ceiling, which may allow particles, such as dust, debris, residue, or other contaminants, to enter the sterile field.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Lots

07613327296167

Skus

CH00000001

Other

RA RA-81701

Model Numbers

CH00000001

Recall Numbers

81701

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - 07613327296167, Model or catalogue number - CH00000001

Risk / hazard

Stryker has discovered that certain lots of the soffit ring may not adequately support the weight of the ceiling cover. As a result, the cover may shift or slide down the ceiling tube. Movement of the ceiling cover can expose the area above the ceiling, which may allow particles, such as dust, debris, residue, or other contaminants, to enter the sterile field.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Stryker Communications

Manufacturer

Not provided

Phone

Not provided

Brand: Stryker Communications.

Source & status

Source

Health Canada

Tier

beta

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

81701

Classification

Type II

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Harmonyair a-series surgical lighting system and alyon surgical lighting system

Medium riskHealth Canada21 May 2026
CA
Recall image placeholder

Harmonyair a-series surgical lighting system

Medium riskHealth Canada12 May 2026
CA
Recall image placeholder

Vectorflow antegrade catheterization product

Medium riskHealth Canada23 Apr 2026
CA
Recall image placeholder

I‑stat eg7+ cartridge

Critical riskHealth Canada15 Apr 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 5 Mar 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.