No image published for this recall

No image published for this recall

Cardiohelp-i

Official notice

13 Oct 2023

Last checked

13 Oct 2023

High riskHealth Canada

An official Health Canada notice concerns Cardiohelp-i. During a complaint investigation it was reported that the rotation of the Cardiohelp Emergency Drive hand crank was blocked or impaired. In course of internal investigations, it was found that the safety washer causes this issue. The safety washer was originally introduced to the Cardiohelp Emergency Drive in order to lengthen the shaft of the emergency drive by the thickness of the safety washer reducing the axial pressure on the circlip, safety washer and bearing. (serial # specific) Recall start date: October 5, 2023 Published 2023-10-13.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74416

Official notice

2023-10-13

The hazard

During a complaint investigation it was reported that the rotation of the Cardiohelp Emergency Drive hand crank was blocked or impaired. In course of internal investigations, it was found that the safety washer causes this issue. The safety washer was originally introduced to the Cardiohelp Emergency Drive in order to lengthen the shaft of the emergency drive by the thickness of the safety washer reducing the axial pressure on the circlip, safety washer and bearing. (serial # specific) Recall start date: October 5, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

74416

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Cardiohelp-I, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 70104.8012

Risk / hazard

During a complaint investigation it was reported that the rotation of the Cardiohelp Emergency Drive hand crank was blocked or impaired. In course of internal investigations, it was found that the safety washer causes this issue. The safety washer was originally introduced to the Cardiohelp Emergency Drive in order to lengthen the shaft of the emergency drive by the thickness of the safety washer reducing the axial pressure on the circlip, safety washer and bearing. (serial # specific) Recall start date: October 5, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Maquet Cardiopulmonary, Cardiohelp-i.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type I

App estimate

high

Recall details

Recall number

74416

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Heater unit hu35

Medium riskHealth Canada18 Sept 2025
CA
No image published for this recall

100925A0 universal remote control

Medium riskHealth Canada13 Nov 2024
CA
No image published for this recall

Meera operating tables

Medium riskHealth Canada19 Sept 2024
CA
No image published for this recall

Cardiohelp-i

Medium riskHealth Canada8 Apr 2024
CA
No image published for this recall

Maquet hls sets

Medium riskHealth Canada9 Jan 2024
CA
No image published for this recall

Cardiohelp-i

Medium riskHealth Canada4 Dec 2023
CA
No image published for this recall

Cardiohelp-i

Medium riskHealth Canada11 Sept 2023
CA
No image published for this recall

Hls set advanced bioline coating

Medium riskHealth Canada25 Jul 2022
CA
No image published for this recall

Hls set advanced 5.0 and 7.0 with bioline coating

Medium riskHealth Canada25 Jul 2022
CA
No image published for this recall

Cardiohelp-i

Medium riskHealth Canada28 Feb 2022
CA
No image published for this recall

Heater unit hu35

Medium riskHealth Canada30 May 2022
CA
No image published for this recall

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Medium riskHealth Canada18 Jun 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 13 Oct 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.