No image published for this recall
Cardiohelp-i
Official notice
11 Sept 2023
Last checked
11 Sept 2023
An official Health Canada notice concerns Cardiohelp-i. It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force. Insufficient securement of the slide rails to the lateral aspects to Cardiohelp could provide an opportunity of both/either the transport guard and/or the emergency drive (with the holder) to detach from Cardiohelp. This issue is serial # specific. Recall Start Date: August 29, 2023 Published 2023-09-11.
What to do now
Contact the manufacturer if you require additional information.
Recall number
74247
Official notice
2023-09-11
The hazard
The hazard
It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force. Insufficient securement of the slide rails to the lateral aspects to Cardiohelp could provide an opportunity of both/either the transport guard and/or the emergency drive (with the holder) to detach from Cardiohelp. This issue is serial # specific. Recall Start Date: August 29, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Maquet Cardiopulmonary, Cardiohelp-i.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Heater unit hu35

Cardiohelp-i

Maquet hls sets

Cardiohelp-i

Cardiohelp-i

Hls set advanced 5.0 and 7.0 with bioline coating

Cardiohelp-i

100925A0 universal remote control

Meera operating tables

Hls set advanced bioline coating

Heater unit hu35
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 11 Sept 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.