No image published for this recall

No image published for this recall

Amplatzer steerable delivery sheath

Official notice

19 Nov 2021

Last checked

19 Nov 2021

Medium riskHealth Canada

An official Health Canada notice concerns Amplatzer steerable delivery sheath. The Hemostasis valve in the closed position is not fully sealed and additional steps are required to prevent air ingress. Abbott is providing instructions (including update to the product instructions for use [IFU] and additional product-related customer training) to reduce the potential for air ingress. Recall start date: 2021-11-09 Published 2021-11-19.

What to do now

Contact the manufacturer if you require additional information

Recall number

63641

Official notice

2021-11-19

The hazard

The Hemostasis valve in the closed position is not fully sealed and additional steps are required to prevent air ingress. Abbott is providing instructions (including update to the product instructions for use [IFU] and additional product-related customer training) to reduce the potential for air ingress. Recall start date: 2021-11-09

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

63641

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected Products - AMPLATZER STEERABLE DELIVERY SHEATH, Lot or serial number - Not applicable, Model or catalog number - ASDS-14F-075

Risk / hazard

The Hemostasis valve in the closed position is not fully sealed and additional steps are required to prevent air ingress. Abbott is providing instructions (including update to the product instructions for use [IFU] and additional product-related customer training) to reduce the potential for air ingress. Recall start date: 2021-11-09

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Abbott Medical, Amplatzer Steerable Delivery Sheath.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

63641

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Abbotts proclaim™ scs proclaim™ drg infinity™ dbs implantable pulse generators

Medium riskHealth Canada30 Apr 2026
CA
No image published for this recall

Centrimag™ adult ecmo kit and blood pump

Critical riskHealth Canada22 Oct 2025
CA
No image published for this recall

Heartmate 3™ system controller

High riskHealth Canada17 Oct 2025
CA
No image published for this recall

Mobile power unit

Medium riskHealth Canada9 Jul 2025
CA
No image published for this recall

Amplatzer torqvue delivery system

Medium riskHealth Canada26 Mar 2025
CA
No image published for this recall

Eterna™ implantable pulse generator

Medium riskHealth Canada15 Oct 2025
CA
No image published for this recall

Tactiflex™ ablation catheter, sensor enabled™

Medium riskHealth Canada17 Sept 2025
CA
No image published for this recall

Cardiomems patient electronics system

Medium riskHealth Canada18 Oct 2024
CA
No image published for this recall

Aveir™ leadless pacemaker

Medium riskHealth Canada19 Aug 2024
CA
No image published for this recall

Heartmate 3 lvas implant kit and heartmate 3 system controller

Medium riskHealth Canada30 Jul 2024
CA
No image published for this recall

Proclaim implantable pulse generators

Medium riskHealth Canada4 Jul 2024
CA
No image published for this recall

Infinity 5 and 7 implantable pulse generator

Medium riskHealth Canada3 Jul 2024

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 19 Nov 2021. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.