Amplatzer steerable delivery sheath
Official notice
19 Nov 2021
Last checked
19 Nov 2021
An official Health Canada notice concerns Amplatzer steerable delivery sheath. The Hemostasis valve in the closed position is not fully sealed and additional steps are required to prevent air ingress. Abbott is providing instructions (including update to the product instructions for use [IFU] and additional product-related customer training) to reduce the potential for air ingress. Recall start date: 2021-11-09 Published 2021-11-19.
What to do now
Contact the manufacturer if you require additional information
Recall number
63641
Official notice
2021-11-19
The hazard
The hazard
The Hemostasis valve in the closed position is not fully sealed and additional steps are required to prevent air ingress. Abbott is providing instructions (including update to the product instructions for use [IFU] and additional product-related customer training) to reduce the potential for air ingress. Recall start date: 2021-11-09
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Abbott Medical, Amplatzer Steerable Delivery Sheath.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Abbotts proclaim™ scs proclaim™ drg infinity™ dbs implantable pulse generators

Centrimag™ adult ecmo kit and blood pump

Heartmate 3™ system controller

Mobile power unit

Amplatzer torqvue delivery system

Eterna™ implantable pulse generator

Tactiflex™ ablation catheter, sensor enabled™

Cardiomems patient electronics system

Aveir™ leadless pacemaker

Heartmate 3 lvas implant kit and heartmate 3 system controller

Proclaim implantable pulse generators

Infinity 5 and 7 implantable pulse generator
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 19 Nov 2021. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.