No image published for this recall
Ambu aview 2 advance
Official notice
14 Feb 2024
Last checked
14 Feb 2024
An official Health Canada notice concerns Ambu aview 2 advance. Ambu have received input from market where Ambu® Aview 2 advance was dropped and upon impact hitting the floor the lithium-ion battery in the device short-circuited and combusted whereby smoke and flames was seen. No harm to patient, user or third party has been reported in any of the cases. Recall start date: August 13, 2023 Published 2024-02-14.
What to do now
Contact the manufacturer if you require additional information.
Recall number
75068
Official notice
2024-02-14
The hazard
The hazard
Ambu have received input from market where Ambu® Aview 2 advance was dropped and upon impact hitting the floor the lithium-ion battery in the device short-circuited and combusted whereby smoke and flames was seen. No harm to patient, user or third party has been reported in any of the cases. Recall start date: August 13, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Ambu A S, Ambu Aview 2 Advance.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Ambu® spur®ii adult and pediatric resuscitator

Ambu® vivasight™ 2 dlt

Ambu aview 2 advance

Ambu vivasight 2 dlt kit

Intramedullary bone saw blade assembly

Penumbra system red 68 reperfusion catheter

Ge healthcare integrated and stand-alone resuscitation systems

Unauthorized soft-shelled hyperbaric chambers may pose serious health risks 2026-06-16

Giraffe™ omnibed™ carestation™

Relay®pro thoracic stent-graft system

Hamilton medical ag breathing set mri

Atlan a350
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 14 Feb 2024. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.