No image published for this recall

No image published for this recall

Ambu aview 2 advance

Official notice

14 Feb 2024

Last checked

14 Feb 2024

Medium riskHealth Canada

An official Health Canada notice concerns Ambu aview 2 advance. Ambu have received input from market where Ambu® Aview 2 advance was dropped and upon impact hitting the floor the lithium-ion battery in the device short-circuited and combusted whereby smoke and flames was seen. No harm to patient, user or third party has been reported in any of the cases. Recall start date: August 13, 2023 Published 2024-02-14.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75068

Official notice

2024-02-14

The hazard

Ambu have received input from market where Ambu® Aview 2 advance was dropped and upon impact hitting the floor the lithium-ion battery in the device short-circuited and combusted whereby smoke and flames was seen. No harm to patient, user or third party has been reported in any of the cases. Recall start date: August 13, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75068

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Ambu Aview 2 Advance, Lot or serial number - 054 to 056 & 058 to 059, Model or catalogue number - 405011000

Risk / hazard

Ambu have received input from market where Ambu® Aview 2 advance was dropped and upon impact hitting the floor the lithium-ion battery in the device short-circuited and combusted whereby smoke and flames was seen. No harm to patient, user or third party has been reported in any of the cases. Recall start date: August 13, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Ambu A S, Ambu Aview 2 Advance.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

75068

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Ambu® spur®ii adult and pediatric resuscitator

Critical riskHealth Canada23 Jul 2025
CA
No image published for this recall

Ambu® vivasight™ 2 dlt

Critical riskHealth Canada2 Oct 2024
CA
No image published for this recall

Ambu aview 2 advance

High riskHealth Canada5 Dec 2023
CA
No image published for this recall

Ambu vivasight 2 dlt kit

Critical riskHealth Canada20 Jun 2022
CA
No image published for this recall

Intramedullary bone saw blade assembly

Medium riskHealth Canada3 Jul 2026
CA
No image published for this recall

Penumbra system red 68 reperfusion catheter

Critical riskHealth Canada22 Jun 2026
CA
No image published for this recall

Ge healthcare integrated and stand-alone resuscitation systems

Medium riskHealth Canada16 Jun 2026
CA
No image published for this recall

Unauthorized soft-shelled hyperbaric chambers may pose serious health risks 2026-06-16

Critical riskHealth Canada16 Jun 2026
CA
No image published for this recall

Giraffe™ omnibed™ carestation™

Medium riskHealth Canada4 Jun 2026
CA
No image published for this recall

Relay®pro thoracic stent-graft system

Critical riskHealth Canada29 May 2026
CA
No image published for this recall

Hamilton medical ag breathing set mri

Medium riskHealth Canada28 May 2026
CA
No image published for this recall

Atlan a350

High riskHealth Canada4 May 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 14 Feb 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.