Image not available in RecallScope.
VERIFY Biological Integrator Challenge Pack Model Catalog Number: LCB007
Official notice
14 Oct 2026
Last checked
20 Oct 2026
An official FDA notice concerns VERIFY Biological Integrator Challenge Pack Model Catalog Number: LCB007. Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction Published 2026-10-14.
What to do now
Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.
Units affected
409
Recall number
Z-0071-2027
Official notice
2026-10-14
The hazard
The hazard
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Lots
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Steris. Contact: Steris Corporation.
Source & status
Source & status
Recall details
Recall details
Recall number
Event id
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.