No image published for this recall

No image published for this recall

700HPM High Pressure Calibration Manifold

Official notice

31 Jul 2024

Last checked

31 Jul 2024

Critical riskOPSS

An official OPSS notice concerns 700HPM High Pressure Calibration Manifold. Affected models present a serious risk of injuries as during manual pressurisation of the equipment, the plunger nut threads of the product may yield, releasing the plunger assembly and allowing it to eject at high velocity. Published 2024-07-31.

What to do now

Fluke UK Ltd has stopped shipping the affected manifold and recalled the product from end users. Further information is available on the Fluke website at the following link: https://www.fluke.com/en-gb/support/safetynotices/fluke-700hxx-recall.

Official notice

2024-07-31

The hazard

Affected models present a serious risk of injuries as during manual pressurisation of the equipment, the plunger nut threads of the product may yield, releasing the plunger assembly and allowing it to eject at high velocity.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Raw Text

pressure-equipment-vessels

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

700HPM High Pressure Calibration Manifold.

Risk / hazard

Affected models present a serious risk of injuries as during manual pressurisation of the equipment, the plunger nut threads of the product may yield, releasing the plunger assembly and allowing it to eject at high velocity.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

GB · United Kingdom

Affected area 1

GB · United Kingdom

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Fluke. Contact: Fluke.

Source & status

Source

Tier

beta

Category

HOME

Classification

product-recall — serious

App estimate

critical

Recall details

Classification

product-recall — serious

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 31 Jul 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.