Zenition 70 and 50

Official notice

10 Oct 2024

Last checked

10 Oct 2024

Medium riskHealth Canada

An official Health Canada notice concerns Zenition 70 and 50. During the system start up or during an ongoing clinical procedure, due to fluctuations in the hospital main power, the fuse can blow thereby causing loss of power to the generator and height positioning module. This results in a system not being available for clinical use. As the system is not available for clinical use, it can result in a delay or abortion of the clinical procedure. Recall Start Date: September 30, 2024 Published 2024-10-10.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76314

Official notice

2024-10-10

The hazard

During the system start up or during an ongoing clinical procedure, due to fluctuations in the hospital main power, the fuse can blow thereby causing loss of power to the generator and height positioning module. This results in a system not being available for clinical use. As the system is not available for clinical use, it can result in a delay or abortion of the clinical procedure. Recall Start Date: September 30, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

76314

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Zenition 70, Lot or serial number - Not applicable., Model or catalog number - 718133

Product 2

Affected products - Zenition 50, Lot or serial number - Not applicable., Model or catalog number - 718096

Risk / hazard

During the system start up or during an ongoing clinical procedure, due to fluctuations in the hospital main power, the fuse can blow thereby causing loss of power to the generator and height positioning module. This results in a system not being available for clinical use. As the system is not available for clinical use, it can result in a delay or abortion of the clinical procedure. Recall Start Date: September 30, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips, Zenition 70 And 50.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

76314

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 10 Oct 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.