No image published for this recall
Wireless recharger
Official notice
2 Dec 2021
Last checked
2 Dec 2021
An official Health Canada notice concerns Wireless recharger. Medtronic has identified through a complaint assessment, that the Medtronic WR9200 recharger, contained within the recharger kit RS6200, have had occurrences where the WR device becomes unresponsive. This issue may occur when the WR is fully depleted and the WR is placed on the dock and quickly removed within 20 seconds, interrupting the initialization routine. The recharger patient user guide instructs the user to place the WR on the dock and charge to full before first use. When the wireless recharger becomes unresponsive, the wireless recharger will not respond to a button press, cannot pair with the recharger app, and will cause the recharger battery indicator to continuously flash when placed on the battery dock. In the event of an unresponsive wireless recharger, the user will not be able to charge their implanted Neurostimulator, until a replacement WR device is received. The unresponsive state of the wireless recharger may lead to unanticipated implanted Neurostimulator INS battery depletion. For the Activa INS, if this unanticipated INS battery depletion (over-discharge state) occurs three times, the battery cannot be recharged and will require surgical intervention. This issue can be prevented by adhering to the instructions outlined in the recharger patient user guide. Recall start date: 2021-11-10 Published 2021-12-02.
What to do now
Contact the manufacturer if you require additional information.
Recall number
63673
Official notice
2021-12-02
The hazard
The hazard
Medtronic has identified through a complaint assessment, that the Medtronic WR9200 recharger, contained within the recharger kit RS6200, have had occurrences where the WR device becomes unresponsive. This issue may occur when the WR is fully depleted and the WR is placed on the dock and quickly removed within 20 seconds, interrupting the initialization routine. The recharger patient user guide instructs the user to place the WR on the dock and charge to full before first use. When the wireless recharger becomes unresponsive, the wireless recharger will not respond to a button press, cannot pair with the recharger app, and will cause the recharger battery indicator to continuously flash when placed on the battery dock. In the event of an unresponsive wireless recharger, the user will not be able to charge their implanted Neurostimulator, until a replacement WR device is received. The unresponsive state of the wireless recharger may lead to unanticipated implanted Neurostimulator INS battery depletion. For the Activa INS, if this unanticipated INS battery depletion (over-discharge state) occurs three times, the battery cannot be recharged and will require surgical intervention. This issue can be prevented by adhering to the instructions outlined in the recharger patient user guide. Recall start date: 2021-11-10
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Medtronic, Wireless Recharger.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 2 Dec 2021. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.