Vns therapy sentiva generators

Official notice

5 Feb 2025

Last checked

5 Feb 2025

Medium riskHealth Canada

An official Health Canada notice concerns Vns therapy sentiva generators. Complaints have been received where the SenTiva generator is reporting a loss of therapy or inability to make therapy changes with the VNS therapy programming system. A reed switch malfunction, resulting in the reed switch sticking in a closed position has been identified as the cause of these device events. Recall start date: January 10, 2025 Published 2025-02-05.

What to do now

Customers to be notified. Instructions to follow on response.

Recall number

76909

Official notice

2025-02-05

The hazard

Complaints have been received where the SenTiva generator is reporting a loss of therapy or inability to make therapy changes with the VNS therapy programming system. A reed switch malfunction, resulting in the reed switch sticking in a closed position has been identified as the cause of these device events. Recall start date: January 10, 2025

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

76909

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - VNS Therapy SenTiva Generators, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 1000

Risk / hazard

Complaints have been received where the SenTiva generator is reporting a loss of therapy or inability to make therapy changes with the VNS therapy programming system. A reed switch malfunction, resulting in the reed switch sticking in a closed position has been identified as the cause of these device events. Recall start date: January 10, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Livanova, Vns Therapy Sentiva Generators.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

76909

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 5 Feb 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.